20

If the overfill security system it’s activated, don’t proceede with the liquid aspiration. If the overfill security 

system doesn’t work there are two cases:

1

st

  case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter 

who avoids the liquid penetration inside the device. 

2

nd

 case – If both the security system and the bacteriological filtrer do not work, there is the possibility that 

liquid has leaked inside the device, in this case return the device to GIMA technical service. 

GIMA S.p.A. will provide upon request electric diagrams, components list, descriptions, setting in-

structions and any other information that can help the technical assistance staff for product repair.

 BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, 

PLEASE  CONTACT  GIMA  TECHNICAL  SERVICE.  GIMA  DOES  NOT  GIVE  GUARANTEE  IF  INSTRU-

MENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE TAMPERED. 

INSTRUCTION FOR USE

•  The device must be checked before each use in order to identify any operating faults and/or damage due 

to transport and/or storage.

•  The work surface must be flat and stable in order that the control panel can be reached and so that the 

vacuum gauge, vase and antibacterial filter are clearly visible.

•  It is recommended not to hold the device in your hand and/or avoid prolonged contact with the body of the 

device.

WARNING: For correct use, position the suction unit on a flat and stable surface, thus maximising the 

available volume of the vase and improving the efficiency of the overflow device.

Operation with AC/DC power pack:

•  Connect the short silicone tube with the antibacterial filter to the suction port. The other tube, connected to 

the filter on one end, must be connected to the spout on the vase lid with the float assembled inside (over-

flow device). The overflow device is triggered (the float closes off the internal lid fitting) when the maximum 

volume (90% of the effective vase volume) is reached, thus preventing the liquid from penetrating the inside 

of the machine. The device must be used on a flat, horizontal surface.

•  Connect the long silicone tube to the free spout on the lid; the free end of the tube must be connected to 

the conical fitting for the probe coupling, to which the suction probe must then be connected.

•  Connect the universal power pack to the device using the dedicated connector and insert the power cord 

plug into the socket. To start the treatment, press the switch into position I to turn the device on

•  Set the desired vacuum level (Bar / kPa) through the vacuum regulator. Turn the knob in a clockwise direc-

tion to increase the vacuum level: these values can be read on the “vacuum gauge”.

•  To suspend and/or end the treatment, press the switch again and pull the plug out of the socket

•  To mitigate the formation of foam inside the vase, unscrew and remove the lid from the vase, and fill the 

latter with 1/3 water (to facilitate cleaning operations and speed up depressurisation during operation), then 

screw the lid back onto the vase.

•  Remove the accessories and proceed with cleaning operations.

•  At the end of each use place the device back in the box, protected against dust

WARNING: The power cord plug is the element of separation from the electrical mains, even if the device is 

equipped with an on/off button. Once the device is in use, the power plug must remain accessible to allow 

another method of disconnection from the electrical mains.

ENGLISH

10. Noisy

Technical internal problem

Contact the technical service

Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 

- 8 - 9 - 10

None  of  the  remedies  has  achieved 

the desired results

Contact  the  seller  or  GIMA  After-sales  Assis-

tance Service

Summary of Contents for 28243

Page 1: ...RADOR SUPER VEGA BATTERY ASPIRADOR SUPER VEGA BATTERY SSAK AKUMULATOROWY SUPER VEGA BATTERY ASPIRATOR SUPER VEGA BATTERY SUPER VEGA BATTERY M28243 M Rev 2 09 20 PROFESSIONAL MEDICAL PRODUCTS Gima S p...

Page 2: ...ilst the plug is still connected disconnect the mains switch remove the plug from the power supply and contact the GIMA technical service department Do not attempt to make the device work before it ha...

Page 3: ...in Home Care Keep all accessories of the device out of reach of children under 36 months of age since they contain small parts that may be swallowed 15 Do not leave the device unattended in places ac...

Page 4: ...rotecting patient and machine against infections cross It blocks the passage of the liquids that come into contact with it Always replace it if yes suspected to be contaminated and or wetted or discol...

Page 5: ...part of the container from secretions under cold running water and then clean every single part in hot water temperature not exceeding 60 C Once again carefully wash each single part using if necessar...

Page 6: ...order to replace it consult the technical staff authorised by the manufacturer USE ONLY THE RECOMMENDED BATTERIES FROM GIMA THE USE OF OTHER BATTERIES ARE NOT RECOMMENDED AND INVOLVING THE CANCELLATIO...

Page 7: ...nd must be connected to the spout on the vase lid with the float assembled inside over flow device The overflow device is triggered the float closes off the internal lid fitting when the maximum volum...

Page 8: ...hicle before the cigarette lighter cable Press the switch to start suction Press the switch to the I position to turn it on WARNING Only use the originally supplied or recommended replacement cigarett...

Page 9: ...lose proximity with adjacent to or on top of the medical device If such use is necessary and unavoidable special precautions should be taken so that the electro medical device functions properly in it...

Page 10: ...The SUPER VEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below The customers or the user of the SUPER VEGA BATTERY SUCTION UNIT should make sure that it s u...

Page 11: ...rmined by an elec tro magnetic study of the sitea could be lower than the level of conformity of each frequency interval b It is possible to check for interference in proximity to devices identified b...

Page 12: ...ransmitters with a maximum nominal output power not shown above the recommended separation distance in metres m can be calculated using the equation applicable to the transmitter frequency where P is...

Page 13: ...rectly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the...

Reviews: