 DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER 

PERIODICAL MAINTENANCE CHECKS

The SUPER VEGA BATTERY SUCTION UNIT suction equipment does not need maintenance or lubrication.

It is, however to inspect the unit before each use. With regard to training, given the information contained in 

the user manual and since it is easy to understand the said device, it doesn’t appear to be necessary. Un-

pack the instrument and always check integrity of plastic parts and AC/DC switching adapter , feeding cable, 

they might have been damaged during previous use. Connect cable to electrical network and turn switch on. 

Close the aspirator outlet with your finger and with suction regulator in maximum vacuum position check that 

the vacuum indicators reaches -75kPa (-0.75 bar) minimum (internal battery). Rotate the knob from right to

left and check the aspiration regulating control. The vacuum indicator should go down -25kPa (-0.25 bar). 

Verify that loud noises are not present, these can indicate wrong functioning. The device is protected by a 

safety fuse (F 10A L 250V) situated in the cigarette lighter cable. When replacing, always check the type and 

value as indicated.

Internally, the device is protected (see electrical specifications) by two fuses F1, F2 (T 15A L 125V) that can-

not be reached from the outside. Therefore, contact the manufacturer to request the assistance of an author-

ized and qualified technician when they need to be replaced. If it’s replaced make sure that its replacement is 

always the same type and value, as indicated.

The device is made up of a lead battery which cannot be accessed by outside. In order to replace it, consult 

the technical staff authorised by the manufacturer.

 USE ONLY THE RECOMMENDED BATTERIES FROM GIMA. THE USE OF OTHER BATTERIES 

ARE NOT RECOMMENDED AND INVOLVING THE CANCELLATION OF WARRANTY

In the event that the service personnel has to replace the internal battery, pay special attention to the polarity 

of the same component. The + / - polarities are indicated directly on the battery.

ENGLISH

Fault type

Cause

Solution

1. Red light on

Battery run down

Hook up the power cord to the electricity mains, 

positioning the equipment power switch on 0.

2. No light

Defective AC/DC adapter or technical 

internal problem

Contact the technical service.

3. No aspiration

Jar Cap badly screwed down

Unscrewed the cap, then rescrew it correctly

4. No aspiration

Lid seal not in its seat

Unscrew the cap and insert the seal properly in 

its seat

5. The float doesn’t close

If  the  cap  has  been  washed,  ensure 

that the float is not partially detached

Insert the float into it’s place

6. The float doesn’t close

The float it’s covered by dirty material Unscrewed the cap, leave the and put in on au-

toclave

7. Low suction

Foam inside the jar

Fill the jar to 1/3 full of ordinary water

8. No aspiration due 

to flow leakage of mucus

Filter blocked

Replace filter

9. The Vacuum power 

on the patient side is either 

very low or absent

•  Vacuum regulator set to minimum

•  Protection filter blocked or damaged

•  Connection tubes blocked, kinked or 

disconnected

•  Shut-off valve blocked or damaged

•  Pump motor damaged

•  Turn  the  vacuum  regulator  clockwise  and 

check the value of the vacuum on the gauge

•  Replace the filter

•  Replace or reconnect the tubes, check the jar 

connections

•  Empty the jar, or disconnect the tube from the 

jar and unblock the shut-off valve. The unit twill 

only work in the upright position

•  Contact the technical service

Summary of Contents for 28243

Page 1: ...RADOR SUPER VEGA BATTERY ASPIRADOR SUPER VEGA BATTERY SSAK AKUMULATOROWY SUPER VEGA BATTERY ASPIRATOR SUPER VEGA BATTERY SUPER VEGA BATTERY M28243 M Rev 2 09 20 PROFESSIONAL MEDICAL PRODUCTS Gima S p...

Page 2: ...ilst the plug is still connected disconnect the mains switch remove the plug from the power supply and contact the GIMA technical service department Do not attempt to make the device work before it ha...

Page 3: ...in Home Care Keep all accessories of the device out of reach of children under 36 months of age since they contain small parts that may be swallowed 15 Do not leave the device unattended in places ac...

Page 4: ...rotecting patient and machine against infections cross It blocks the passage of the liquids that come into contact with it Always replace it if yes suspected to be contaminated and or wetted or discol...

Page 5: ...part of the container from secretions under cold running water and then clean every single part in hot water temperature not exceeding 60 C Once again carefully wash each single part using if necessar...

Page 6: ...order to replace it consult the technical staff authorised by the manufacturer USE ONLY THE RECOMMENDED BATTERIES FROM GIMA THE USE OF OTHER BATTERIES ARE NOT RECOMMENDED AND INVOLVING THE CANCELLATIO...

Page 7: ...nd must be connected to the spout on the vase lid with the float assembled inside over flow device The overflow device is triggered the float closes off the internal lid fitting when the maximum volum...

Page 8: ...hicle before the cigarette lighter cable Press the switch to start suction Press the switch to the I position to turn it on WARNING Only use the originally supplied or recommended replacement cigarett...

Page 9: ...lose proximity with adjacent to or on top of the medical device If such use is necessary and unavoidable special precautions should be taken so that the electro medical device functions properly in it...

Page 10: ...The SUPER VEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below The customers or the user of the SUPER VEGA BATTERY SUCTION UNIT should make sure that it s u...

Page 11: ...rmined by an elec tro magnetic study of the sitea could be lower than the level of conformity of each frequency interval b It is possible to check for interference in proximity to devices identified b...

Page 12: ...ransmitters with a maximum nominal output power not shown above the recommended separation distance in metres m can be calculated using the equation applicable to the transmitter frequency where P is...

Page 13: ...rectly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the...

Reviews: