Recommended separation distance between portable and mobile radio-communication devices and the monitor
The  SUPER  VEGA  BATTERY  SUCTION  UNIT  surgical  aspirator  is  intended  to  operate  in  an  electro-magnetic  envi-

ronment where RF irradiated interferences are under control. The client or operator of the SUPER VEGA BATTERY 

SUCTION UNIT device can help prevent electro-magnetic interference by keeping a minimum distance between the 

portable and mobile RF communication devices (transmitters) and the SUPER VEGA BATTERY SUCTION UNIT device, 

as recommended below, in relation to the radio-communication maximum output power.
Maximum nominal

output power of the

Transmitter W

Separation distance from the frequency transmitter (m)

150KHz to 80MHz

80MHz to 800MHz

800MHz to 2,7GHz 

0,01

0,12

0,12

0,23

0,1

0,38

0,38

0,73

1

1,2

1,2

2,3

10

3,8

3,8

7,3

100

12

12

23

For transmitters with a maximum nominal output power not shown above, the recommended separation distance in metres 

(m) can be calculated using the equation applicable to the transmitter frequency, where 

P

 is the maximum nominal output 

power of the transmitter in Watt (W) depending on the transmitter’s manufacturer.

Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied

Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the

absorption and by the reflection from buildings, objects and people

3,5

 V

1

√P

d

=

12

 E

1

√P

d=

 

23

 E

1

√P

d=

Caution: read instructions (warnings) carefully

Follow instructions for use

Keep in a cool, dry place

Keep away from sunlight

Manufacturer  

Date of manufacture

Product code

Lot number 

Medical Device complies 

with Directive 93/42/EEC 

Type BF applied part

WEEE disposal

Class II applied

Serial number

Temperature limit 

Battery

Atmospheric Pressure limit

Direct current

~

Alternating current

IP21

Covering Protection rate 

Hz

Mains frequency

ON / OFF

Humidity limit 

SYMBOLS

Summary of Contents for 28243

Page 1: ...RADOR SUPER VEGA BATTERY ASPIRADOR SUPER VEGA BATTERY SSAK AKUMULATOROWY SUPER VEGA BATTERY ASPIRATOR SUPER VEGA BATTERY SUPER VEGA BATTERY M28243 M Rev 2 09 20 PROFESSIONAL MEDICAL PRODUCTS Gima S p...

Page 2: ...ilst the plug is still connected disconnect the mains switch remove the plug from the power supply and contact the GIMA technical service department Do not attempt to make the device work before it ha...

Page 3: ...in Home Care Keep all accessories of the device out of reach of children under 36 months of age since they contain small parts that may be swallowed 15 Do not leave the device unattended in places ac...

Page 4: ...rotecting patient and machine against infections cross It blocks the passage of the liquids that come into contact with it Always replace it if yes suspected to be contaminated and or wetted or discol...

Page 5: ...part of the container from secretions under cold running water and then clean every single part in hot water temperature not exceeding 60 C Once again carefully wash each single part using if necessar...

Page 6: ...order to replace it consult the technical staff authorised by the manufacturer USE ONLY THE RECOMMENDED BATTERIES FROM GIMA THE USE OF OTHER BATTERIES ARE NOT RECOMMENDED AND INVOLVING THE CANCELLATIO...

Page 7: ...nd must be connected to the spout on the vase lid with the float assembled inside over flow device The overflow device is triggered the float closes off the internal lid fitting when the maximum volum...

Page 8: ...hicle before the cigarette lighter cable Press the switch to start suction Press the switch to the I position to turn it on WARNING Only use the originally supplied or recommended replacement cigarett...

Page 9: ...lose proximity with adjacent to or on top of the medical device If such use is necessary and unavoidable special precautions should be taken so that the electro medical device functions properly in it...

Page 10: ...The SUPER VEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below The customers or the user of the SUPER VEGA BATTERY SUCTION UNIT should make sure that it s u...

Page 11: ...rmined by an elec tro magnetic study of the sitea could be lower than the level of conformity of each frequency interval b It is possible to check for interference in proximity to devices identified b...

Page 12: ...ransmitters with a maximum nominal output power not shown above the recommended separation distance in metres m can be calculated using the equation applicable to the transmitter frequency where P is...

Page 13: ...rectly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the...

Reviews: