The technical specifications may change without notice
CLEANING OF THE DEVICE
Use a soft dry cloth with not – abrasive and not – solvent detergents. To clean the device external parts always
use a cotton cloth dampened with detergent. Don’t use abrasive or solvent detergents. Before carrying out
any cleaning and / or maintenance operation, disconnect the appliance from the power supply, unplugging it
or turning off the switch on the device.
Particular care should be taken to ensure that the internal parts of the equipment do not get in
touch with liquids. Never clean the equipment under water.
During all clearing operations use protection gloves and apron (if need be, also wear a face mask and glass-
es) to avoid getting in contact with contaminating substances (after each utilization cycle of the machine).
ACCESSORIES SUPPLIES
• Complete aspiration jar 1000ml
• Conical fitting
• Tubes set 6 mm x 10 mm
• Hydrophobic and antibacterial filter
• AC/DC adapter
• Power supply cord for ac/dc adapter
• Cigarette ligther cable
On request, versions with a complete 2000ml jar are also available.
Antibacterial and hydrophobic filter: designed for a single patient with the aim of protecting patient and
machine against infections cross. It blocks the passage of the liquids that come into contact with it. Always
replace it if yes suspected to be contaminated and / or wetted or discolored. If the aspirator is used on patients
in pathological situations not known and where it is not possible to assess any indirect contamination, replace
the filter after each use. The filter is not designed to be decontaminated, disassembled and / or sterilized.
In case the pathology of the patient and / or is knownwhere there is no danger of indirect contamination, we
recommend replacing the filter after each work shift or otherwise every month even if the device is not used.
Suction catheter: Single-use device to be used on a single patient. Do not wash or re-sterilize after use.
Reuse may cause cross infections. Don’t use after lapse of the sell-by date
Insulation Class (when used with an Internal battery)
Internally Powered Equipment
Insulation Class (when used with a car cigarette
lighter cable)
Class II
Size
350 x 210 x 180 mm
Battery Holding Time
80 minutes
Battery Time Charge
240 minutes
Accuracy of Vacuum Indicator
± 5%
Working Condition
Room temperature: 5 ÷ 35°C
Room humidity percentage: 10 ÷ 93 % RH
Atmospheric pressure: 800 ÷ 1060 hPa
Conservation condition and Transport
Room temperature: - 25 ÷70°C
Room humidity percentage: 0 ÷ 93% RH
Atmospheric pressure: 500 ÷ 1060 hPa