background image

A

C

C

ES
SO

R

IES

11

PIPETMAN L®

 

|

 

USER’S GUIDE

3.  Change the tip before aspirating a different liquid, sample, or reagent.
4.  Change the tip if a droplet remains at the end of the tip from the previous pipetting operation.
5.  Each new tip should be pre-rinsed with the liquid to be pipetted.
6.  Liquid should 

never

 enter the tip holder; to prevent this:

 

press and release the push button slowly and smoothly,

 

never turn the pipette upside down,

 

never lay the pipette on its side when there is liquid in the tip.

7.  If you use the same tip with a higher volume, pre-rinse the tip.
8.  For volatile solvents you should saturate the air cushion in your pipette by aspirating and dispensing 

the solvent repeatedly before aspirating the sample.

9.  When the temperature of the liquid is different from the ambient temperature, pre-rinse the tip 

several times before use.

10.  You may remove the tip ejector (refer to 

Change the Tip Ejector

 on page 14) to aspirate from very 

narrow tubes.

11.  After pipetting acids or other corrosive liquids that emit vapors, remove the tip ejector and tip holder, 

rinse with distilled water, dry, and lubricate the piston (refer to 

MAINTENANCE

 on page 14). For the 

P1000L model, use a specific tip holder equipped with a filter to increase the lifetime of the piston 

(refer to 

ACCESSORIES

 on page 11).

12.  Do not pipette liquids with temperatures above 70°C or below 4°C. The pipette can be used between 

+4°C and +40°C but the specifications may vary according to the temperature (refer to the ISO 8655-2 

standard for conditions of use).

Chapter 5

ACCESSORIES

To make pipetting more comfortable and more secure, Gilson has developed several accessories:

SINGLE™

CARROUSEL™

TRIO™

COLORIS™

Universal Multichannel 

Stand™

Corrosion protection kit

Pipette stands allow users to store pipettes vertically 

to avoid the possibility of liquid running back into the 

pipette.

DESCRIPTION 

P/N

CARROUSEL™ pipette stand (7 pipettes)

F161401

TRIO™ stand (3 pipettes)

F161405

Universal multichannel stand*

F161417

Single™ pipette holder

F161406

To identify or personalize your pipette, COLORIS™ clips 

are available:

DESCRIPTION 

P/N

COLORIS™ clips (mixed colors set of 10) 

F161301

COLORIS clips (red, set of 10) 

F161302 

COLORIS clips (yellow, set of 10) 

F161303

COLORIS clips (green, set of 10) 

F161304

COLORIS clips (blue, set of 10) 

F161305

COLORIS clips (white, set of 10) 

F161306

With The JIMMY™, a hands-free microtube opener, you 

can open both snap-cap and screw-cap microtubes.

DESCRIPTION 

P/N

The JIMMY™ (set of 3)

F144983

To protect the piston when pipetting corrosive liquids, 

you can use a specific tip holder and filter for the 

model P1000L:

DESCRIPTION 

P/N

Corrosion Protection Kit  

(tip a bag of 10 filters)

F144570

Corrosion Protection Kit Filter (Bag of 10)

F144571

Corrosion Protection Kit Filter (Bag of 50)

F144572

Universal multichannel stand is for multichannel models only.

Disposable Reagent Reservoirs are ideal for 

dispensing reagent with multichannel pipettes:

DESCRIPTION 

P/N

REAGENT RESERVOIR 25 mL, x100

F267660

REAGENT RESERVOIR 50 mL, x100

F267670

Summary of Contents for FA10001M

Page 1: ...Variable volume single and multichannel models and fixed volume models A UTOCLAVA B L E FULLY VARIABLE VOLUME MODELS User s Guide EN...

Page 2: ...t 13 MAINTENANCE FOR SINGLE AND FIXED MODELS 14 Change the Tip Ejector 14 Change the Tip Holder 14 Service the Piston Assembly 15 Change the Seals 16 MAINTENANCE FOR MULTICHANNEL MODELS 16 CLEANING AN...

Page 3: ...ures The Serial Number is engraved on the body of the pipette It provides unique identification of your pipette and the date of manufacture Example GB58672 The Bar Code on the box and the certi cate o...

Page 4: ...ejector Graduated cover Tip holder Personal label You can identify your pipette with a name tag 1 Pry out the window by inserting a small screwdriver in the access slot 2 Position the name tag in the...

Page 5: ...5 0 012 0 12 0 08 5 0 075 0 030 0 12 0 08 Max 10 0 100 0 040 0 12 0 08 P20L P N FA10003P OR FA10003M Min 2 0 10 0 03 0 20 0 10 10 0 10 0 05 0 20 0 10 Max 20 0 20 0 06 0 20 0 10 P100L P N FA10004P OR F...

Page 6: ...ctor the part number is FA10002P For the same pipette with a stainless steel tip ejector the part number is FA10002M NOTICE INTRODUCTION 6 INTRODUCTION PIPETMAN L FIXED MODELS MODEL VOL L GILSON ISO 8...

Page 7: ...bwheel the new volume selected is locked To obtain maximum accuracy when setting the volume proceed as follows When decreasing the volume setting slowly reach the required setting making sure not to o...

Page 8: ...AN L Fixed is readjusted to compensate for a particular solution the nature of the liquid can be noted on the labels inserted in the slot under the counter Figure 7 Adjust PIPETMAN L Fixed volume sett...

Page 9: ...our Tipacks and Towerpacks Ejector spacer for PIPETMAN L Multichannel 10 L According to the tip used D10 or DL10 you might have to exchange the ejector spacer the broad one is dedicated to D10 and the...

Page 10: ...along the inside surface of the vessel 4 Release the push button smoothly Eject the tip by pressing firmly on the tip ejector button For the multichannel models use a Gilson reagent reservoir refer t...

Page 11: ...C but the speci cations may vary according to the temperature refer to the ISO 8655 2 standard for conditions of use Chapter 5 ACCESSORIES To make pipetting more comfortable and more secure Gilson ha...

Page 12: ...nscrew tip holder Tighten connecting nut Connecting nut is loose Tighten connecting nut Pipette is not precise Tip holder is loose Tighten connecting nut Connecting nut is loose Tighten connecting nut...

Page 13: ...vel goes down there is a leak For the P1000L P5000L and P10mL L Single channel models for the F300L to F5000L Fixed models 1 Fit a PIPETMAN DIAMOND tip 2 For the variable volume models set the pipette...

Page 14: ...emove the tip ejector To refit 1 Push the ejection button 2 Slide the tip ejector along the tip holder 3 Clip the tip ejector on the body of the pipette Change the Tip Holder no tools required 1 Remov...

Page 15: ...miniaturized parts it is best not to disassemble these pipettes yourself 1 Remove the tip ejector refer to Change the Tip Ejector page 14 2 Unscrew the connecting nut by turning it clockwise 3 Carefu...

Page 16: ...he seal lubricate them lightly Please place them in the correct order 6 Reassemble the pipette refer to the gure 18 on page 15 7 Tighten the connecting nut turn counter clockwise 8 Re t the tip ejecto...

Page 17: ...seals must be degreased with isopropanol or ethanol before being immersed in another ultra sonic bath Small round brushes with soft plastic bristles may be used to clean the interior of the tip holder...

Page 18: ...t cloth or lint free tissue soaked with distilled water or sterile water Lower Part Volumetric module The following components only can be immersed in a decontaminant solution connecting nut tip eject...

Page 19: ...SCRIPTION P200L P1000L C D E Service Kit 1st level FA07004 FA07005 A B F Service Kit 2nd level FA07305 FA07306 C D Seal guide O ring 5 sets FA07015 FA07016 P5000L FA10007 AND P10ML L FA10008 LABEL DES...

Page 20: ...F507003 H3 H4 Ejector spacer D10 F807114 F807115 P8X20L FA10009 AND P12X20L FA10010 LABEL DESCRIPTION 8X20 12X20 F3 F4 Tip ejector F507005 F507006 G Ejector lock F507008 F507008 H1 H2 Ejector spacer F...

Page 21: ...F100L FA10024 LABEL DESCRIPTION F20L F25L F50L F100L C D E Service Kit 1st level FA07002 FA07003 A B F1 Service Kit 2nd level FA07069 FA07070 C D Seal guide O ring 5 sets FA07013 FA07014 F200L FA1002...

Page 22: ...void in the event pipettes are altered or modified by any party other than Gilson or its designates Gilson s sole liability under this warranty shall be limited to at Gilson s sole option repair or re...

Page 23: ...Notes 23 PIPETMAN L USER S GUIDE NOTES...

Page 24: ...USA T 608 836 1551 or 800 445 7661 F 608 831 4451 Gilson S A S 19 avenue des Entrepreneurs BP 145 F 95400 Villiers le Bel France T 33 0 1 34 29 50 00 F 33 0 1 34 29 50 20 LT801575 B 08 2017 Gilson Inc...

Reviews: