The data given in the tables conform to the ISO 8655-2 Standard.
With a precise pipetting technique (refer to
Guidelines for Good Pipetting
,
page 10) the P2L model may be used to aspirate volumes as low as
0.1 µL and the P10L model as low as 0.5 µL.
NOTICE
Each single pipette model (except P5000L and P10mL L) has two
different part numbers to identify the kind of tip ejector required. For
a pipette with a plastic tip ejector, the part number is ended by the
letter P, for a pipette with a stainless steel tip ejector, the part number
is ended by the letter M.
Ex: For a P10L model with the plastic tip ejector the part number is
FA10002P. For the same pipette with a stainless steel tip ejector, the
part number is FA10002M
NOTICE
INTR
O
D
U
C
TI
O
N
6
INTRODUCTION
|
PIPETMAN L®
FIXED MODELS
MODEL
VOL
(µL)
GILSON
ISO 8655
SYSTEMATIC
ERROR
RANDOM
ERROR
SYSTEMATIC
ERROR
RANDOM
ERROR
F1L (FA10017)
1
±0.020
≤
0.015
±0.05
≤
0.05
F2L (FA10018)
2
±0.050
≤
0.020
±0.08
≤
0.04
F5L (FA10019)
5
±0.05
≤
0.025
±0.125
≤
0.075
F10L (FA10020)
10
±0.060
≤
0.030
±0.12
≤
0.08
F20L (FA10021)
20
±0.100
≤
0.050
±0.2
≤
0.1
F25L (FA10022)
25
±0.200
≤
0.070
±0.5
≤
0.2
F50L (FA10023)
50
±0.35
≤
0.12
±0.5
≤
0.2
F100L (FA10024)
100
±0.55
≤
0.15
±0.8
≤
0.3
F200L (FA10025)
200
±1.20
≤
0.30
±1.6
≤
0.6
F250L (FA10026)
250
±1.50
≤
0.75
±4
≤
1.5
F300L (FA10027)
300
±2.4
≤
0.5
±4
≤
1.5
F400L (FA10028)
400
±2.4
≤
0.8
±4
≤
1.5
F500L (FA10029)
500
±3.0
≤
0.8
±4
≤
1.5
F1000L (FA10030)
1000
±5.0
≤
1.3
±8
≤
3
F5000L (FA10031)
5000
±20.0
≤
7.0
±40
≤
15