P10L, P20L
P100L, P200L
P5000L
P10mL L
P1000L
Tip holder
Tip ejector
stainless steel
plastic
Connecting
nut
Color coded
push button
Piston
assembly
Thumbwheel
Variable volume single channel models
Fixed models
F10L, F20L
F100L, F200L
F5000L
F1000L
Tip holder
Tip ejector
Connecting
nut
Piston
assembly
Color coded
push button
MA
IN
TE
NAN
CE
F
O
R
S
IN
G
LE
AN
D
F
IX
E
D
M
O
D
E
LS
15
PIPETMAN L®
|
USER’S GUIDE
Figure 18
Piston Assembly (Disassembled)
Service the Piston Assembly
You may remove the piston assembly for cleaning purposes only. If the piston assembly is changed, the
pipette must be adjusted and calibrated in a Gilson-authorized Service Center. As the models P2L and
P10L, F1L, F2L, F5L and F10L contain miniaturized parts, it is best not to disassemble these pipettes
yourself.
1.
Remove the tip ejector (refer to
Change the Tip Ejector
, page 14).
2. Unscrew the connecting nut by turning it clockwise.
3. Carefully separate the lower and upper parts.
4. Remove the piston assembly, O-ring and seal.
5. Leave exposed the piston, clean it with isopropanol or ethanol and lubricate lightly.
For P20L, P100L, P200L, F20L, F25L, F50L, F100L, F200L, F250L
: lubricate only
the useful part of the piston (20 ±5 mm length) and the O-ring.
For P1000L, F300L, F400L, F500L, and F1000L
: lubricate the piston.
For P5000L, P10mL L, and F5000L
: disassemble the seals, lubricate their internal
part and lubricate the piston. Do not lubricate the O-ring.
6. Reassemble the pipette (refer to figure 18).
7. Tighten the connecting nut (turn counter-clockwise).
8. Refit the tip ejector (refer to
Change the Tip Ejector
, page 14).
Figure 19
Hold the piston
assembly