PIPETTING

 

PIPETMAN L®

Graduations

Counterclockwise

Clockwise

To adjust the volume setting to compensate for a 

specific density or viscosity, engage the two hooks 

of the spanner tool inside the two small holes on the 

top of the body, then turn the key slowly:

•  Clockwise

 to decrease the volume; making sure 

not to overshoot the mark.

•  Counterclockwise

 to increase the volume; pass 

the required value by 1/4 turn, then slowly decrease 

the volume to reach the required setting.

Example: When using a PIPETMAN L Fixed 100 to aspirate 

a particularly viscous solution, you may determine 

gravimetrically that the volume delivered is 98.8 µL, and the 

display reads “H”. Therefore, you wish to increase the volume 

dispensed by 1.2 µL. As the interval between each letter on 

the calibration display represents 0.2 µL for the F100L (1/10 

of 2 µL), the calibration tool must be turned 3 units until the 

display reads “B”.

It is possible to alter the volume by more than one full 

turn in the counterclockwise direction. In this case, 

remember to readjust the same number of turns when 

it is necessary to return to the original setting.

Check the new volume gravimetrically. If the volumes 

delivered are still not sufficiently close to the desired 

value, make another slight adjustment in the relevant 

direction. 

Be sure to change tips between volume 

setting adjustments and to pre-rinse each new tip.

When the PIPETMAN L Fixed is readjusted to 

compensate for a particular solution, the nature of 

the liquid can be noted on the labels inserted in the 

slot under the counter. 

Figure 7

Adjust PIPETMAN L Fixed volume setting to 

compensate for a specific density or viscosity

Liquids of high density, vapor 

pressure, or viscosity are best 

pipetted by the MICROMAN 

positive displacement pipette.

NOTICE

Figure 8

Identify the liquid used with the label

Chapter 4

PIPETTING

It is recommended to use PIPETMAN® DIAMOND Tips with the PIPETMAN L for optimum performance. 

These tips are made from pure polypropylene. Plastic tips are for a single application—they must not be 

cleaned for reuse.

Fit the Tips

SINGLE CHANNEL MODELS

To fit a new PIPETMAN DIAMOND tip, push the tip holder into the tip using a slight twisting motion to 

ensure a firm and airtight seal.

For the P2L and P10L models equipped with a stainless steel tip ejector

, a dual-position adapter (plastic) 

is required to fit DL10 tips (long tips) or D10 tips (short tips). The metallic rod of the tip ejector is shaped so 

that the adapter may be clipped to it in either of two positions.

PIPETMAN DIAMOND Tip Compatibility 

for Fixed models

F1L, F2L, F5L, F10L 

 

D10, DL10, DF10, DFL10

F20L, F25L 

 

D200, DF30

F50L, F100L 

 

D200, DF100

F200L 

 

D200, DF200

F250L 

 

D300, DF300

F300L, F400L, F500L, F1000L 

D1000, DF1000

F5000 

 D5000

PIPETMAN DIAMOND Tip Compatibility 

for Single models

P2L, P10L 

D10, DL10, DF10, DFL10

P20L 

D200, DF30

P100L 

D200, DF100

P200L 

D200, DF200, D300, DF300

P1000L 

D1000, D1200, DF1000, DF1200

P5000L D5000

P10mL L 

D10mL

Figure 9

PIPETMAN DIAMOND Tip compatibility charts

Summary of Contents for FA10001M

Page 1: ...Variable volume single and multichannel models and fixed volume models A UTOCLAVA B L E FULLY VARIABLE VOLUME MODELS User s Guide EN...

Page 2: ...t 13 MAINTENANCE FOR SINGLE AND FIXED MODELS 14 Change the Tip Ejector 14 Change the Tip Holder 14 Service the Piston Assembly 15 Change the Seals 16 MAINTENANCE FOR MULTICHANNEL MODELS 16 CLEANING AN...

Page 3: ...ures The Serial Number is engraved on the body of the pipette It provides unique identification of your pipette and the date of manufacture Example GB58672 The Bar Code on the box and the certi cate o...

Page 4: ...ejector Graduated cover Tip holder Personal label You can identify your pipette with a name tag 1 Pry out the window by inserting a small screwdriver in the access slot 2 Position the name tag in the...

Page 5: ...5 0 012 0 12 0 08 5 0 075 0 030 0 12 0 08 Max 10 0 100 0 040 0 12 0 08 P20L P N FA10003P OR FA10003M Min 2 0 10 0 03 0 20 0 10 10 0 10 0 05 0 20 0 10 Max 20 0 20 0 06 0 20 0 10 P100L P N FA10004P OR F...

Page 6: ...ctor the part number is FA10002P For the same pipette with a stainless steel tip ejector the part number is FA10002M NOTICE INTRODUCTION 6 INTRODUCTION PIPETMAN L FIXED MODELS MODEL VOL L GILSON ISO 8...

Page 7: ...bwheel the new volume selected is locked To obtain maximum accuracy when setting the volume proceed as follows When decreasing the volume setting slowly reach the required setting making sure not to o...

Page 8: ...AN L Fixed is readjusted to compensate for a particular solution the nature of the liquid can be noted on the labels inserted in the slot under the counter Figure 7 Adjust PIPETMAN L Fixed volume sett...

Page 9: ...our Tipacks and Towerpacks Ejector spacer for PIPETMAN L Multichannel 10 L According to the tip used D10 or DL10 you might have to exchange the ejector spacer the broad one is dedicated to D10 and the...

Page 10: ...along the inside surface of the vessel 4 Release the push button smoothly Eject the tip by pressing firmly on the tip ejector button For the multichannel models use a Gilson reagent reservoir refer t...

Page 11: ...C but the speci cations may vary according to the temperature refer to the ISO 8655 2 standard for conditions of use Chapter 5 ACCESSORIES To make pipetting more comfortable and more secure Gilson ha...

Page 12: ...nscrew tip holder Tighten connecting nut Connecting nut is loose Tighten connecting nut Pipette is not precise Tip holder is loose Tighten connecting nut Connecting nut is loose Tighten connecting nut...

Page 13: ...vel goes down there is a leak For the P1000L P5000L and P10mL L Single channel models for the F300L to F5000L Fixed models 1 Fit a PIPETMAN DIAMOND tip 2 For the variable volume models set the pipette...

Page 14: ...emove the tip ejector To refit 1 Push the ejection button 2 Slide the tip ejector along the tip holder 3 Clip the tip ejector on the body of the pipette Change the Tip Holder no tools required 1 Remov...

Page 15: ...miniaturized parts it is best not to disassemble these pipettes yourself 1 Remove the tip ejector refer to Change the Tip Ejector page 14 2 Unscrew the connecting nut by turning it clockwise 3 Carefu...

Page 16: ...he seal lubricate them lightly Please place them in the correct order 6 Reassemble the pipette refer to the gure 18 on page 15 7 Tighten the connecting nut turn counter clockwise 8 Re t the tip ejecto...

Page 17: ...seals must be degreased with isopropanol or ethanol before being immersed in another ultra sonic bath Small round brushes with soft plastic bristles may be used to clean the interior of the tip holder...

Page 18: ...t cloth or lint free tissue soaked with distilled water or sterile water Lower Part Volumetric module The following components only can be immersed in a decontaminant solution connecting nut tip eject...

Page 19: ...SCRIPTION P200L P1000L C D E Service Kit 1st level FA07004 FA07005 A B F Service Kit 2nd level FA07305 FA07306 C D Seal guide O ring 5 sets FA07015 FA07016 P5000L FA10007 AND P10ML L FA10008 LABEL DES...

Page 20: ...F507003 H3 H4 Ejector spacer D10 F807114 F807115 P8X20L FA10009 AND P12X20L FA10010 LABEL DESCRIPTION 8X20 12X20 F3 F4 Tip ejector F507005 F507006 G Ejector lock F507008 F507008 H1 H2 Ejector spacer F...

Page 21: ...F100L FA10024 LABEL DESCRIPTION F20L F25L F50L F100L C D E Service Kit 1st level FA07002 FA07003 A B F1 Service Kit 2nd level FA07069 FA07070 C D Seal guide O ring 5 sets FA07013 FA07014 F200L FA1002...

Page 22: ...void in the event pipettes are altered or modified by any party other than Gilson or its designates Gilson s sole liability under this warranty shall be limited to at Gilson s sole option repair or re...

Page 23: ...Notes 23 PIPETMAN L USER S GUIDE NOTES...

Page 24: ...USA T 608 836 1551 or 800 445 7661 F 608 831 4451 Gilson S A S 19 avenue des Entrepreneurs BP 145 F 95400 Villiers le Bel France T 33 0 1 34 29 50 00 F 33 0 1 34 29 50 20 LT801575 B 08 2017 Gilson Inc...

Reviews: