Series 8650A Universal Power Meters
3-2
Manual 31470, Rev. E, April 2001
Clear Device
The interface command CLEAR 713 resets the GPIB and sets the 8650A to its preset condition.
Clear Interface
The interface command ABORT 7 resets the GPIB without resetting the 8650A to its preset condition.
The 8650A will not be addressed after the abort.
Local and Remote Control
The interface command LOCAL 713 places the 8650A into the local control mode.
The interface command REMOTE 713 places the 8650A into the remote control mode. Enter LOCAL
713 to return the instrument to local mode.
The interface command LOCAL LOCKOUT 7 places the 8650A in the local lockout mode. This is a
remote control mode in which all of the 8650A front panel keys are disabled. The GPIB LOCAL
command must be issued to return the 8650A to local mode (disconnecting the GPIB cable will also
return the instrument to local mode).
Sensor Selection and Calibration
Power sensor selection data, specifications, and calibration (local and remote) are contained in
Appendix B of this manual.
3.1.2
Polling
The GPIB supports parallel and serial polling. The example programs below show how to use the
parallel and serial poll capabilities of the 8650A to determine when a requested zeroing operation is
completed.
Parallel Polling
Ppoll_zero
! zero using parallel poll
PRINT entering parallel poll zero routine
PPOLL CONFIGURE 713;8
! configure response on bit zero
OUTPUT 713;CS AEZE
! clear status byte, zero channel A
State=0
! initialize variable
WHILE State 1
! stay here until zero done
State=PPOLL(7)
! read the poll
END WHILE PPOLL UNCONFIGURE 713 ! cancel parallel poll mode
PRINT parallel zero done RETURN
Summary of Contents for 8650A Series
Page 4: ......
Page 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Page 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Page 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Page 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Page 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Page 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Page 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Page 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Page 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...