Remote Operation
Manual 31470, Rev. E, April 2001
3-27
3.2.16
High Speed Measurements
CALCulate:DATA?
CALCulate:MODE
TRIGger:MODE
TRIGger:DELay
TRIGger:COUNt
Measurements in Normal mode are fastest with only one sensor attached. When two sensors are
attached, the Normal mode measurement rate is reduced. This applies for all three major measurement
commands, FETCh, READ, and MEASure. Both of the averaging types, MOVing and REPeat, slow
down when in the Normal mode. MOVing provides faster averaging; the speed is equivalent to REPeat
whenever the Averaging number is 1.
When performing measurements in either Swift or Burst modes, measurement rates are the same with
two sensors as with one sensor. Approximate measuring speeds are listed in Table 1-1.
CALC:DATA?
Syntax
:
CALCulate<channel 1 or 2>:DATA?
Example
:
OUTPUT @Pwr_mtr;CALC1:DATA?
! Query channel 1 & 2 BURSt mode data
Description
: This command interrupts the burst measurement and dumps the measurement buffer. All
empty locations will be filled with -300 before the data dump. The Burst mode remains
active after this command so that a trigger will start a new measurement reference.
The command returns measurement data (while the measurement is completing) during
BURSt mode only. This command is not used in NORMal or SWIFt Modes. Operation is
similar to using FETCh? Both are instrument (not channel) commands. That is, when both
sensors are connected, calibrated, and the channel set to ON, this command will return
power level data for both channels.
Operating Mode Control
The Operating mode is controlled through the CALC#:MODE command. The choices are NORMal,
BURSt, and SWIFt. Sending CALC1:MODE will set the operating mode for the entire instrument, not
just for channel 1. For example, you cannot send CALC1:MODE BURS, and then CALC2:MODE
SWIF to get channel 1 in BURSt mode and channel 2 in SWIFt mode. After the CALC2:MODE SWIF
command, both channels are in SWIFt mode.
CALC:MODE
Syntax
:
CALCulate<channel 1 or 2>:MODE
space<NORMal, BURSt, or SWIFt Mode Selection>
Example
:
OUTPUT @Pwr_mtr;CALC2:MODE SWIF
! Set measurement mode to SWIFt
Description
: This command sets the measurement mode. This is NOT a channel-specific command; it
is an instrument configuration command. Selecting one channel changes both channels to
the desired mode. Use this command to select the level of functionality and performance
for your unique applications. NORMal Mode is full functioned. SWIFt Mode allows individ-
ual data point triggering. BURSt Mode allows the fastest measurement rates.
Summary of Contents for 8650A Series
Page 4: ......
Page 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Page 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Page 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Page 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Page 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Page 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Page 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Page 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Page 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...