Remote Operation
Manual 31470, Rev. E, April 2001
3-41
CALC:MIN
Syntax
:
CALCulate<channel 1 or 2>:MINimum:STATe
space
ON or OFF
Example
:
OUTPUT @Pwr_mtr;CALC1:MIN:STAT ON
! Enable channel 1 minimum tracking
Description
: This command activates the minimum channel value monitor.
Syntax
:
CALCulate<channel 1 or 2>:MINimum [:MAGnitude]?
Example
:
OUTPUT @Pwr_mtr;CALC1:MIN?
! Query channel minimum value in dBm
Response
:
Lowest power reading since CALC2:MIN:STAT ON was sent.
Description
: This command reports the value maintained as the minimum in the 8650A min monitor.