Remote Operation
Manual 31470, Rev. E, April 2001
3-95
3.17.4
Swift Mode
Swift mode is a fast mode which allows for fast continuous data taking and return of each measurement
to the host as it is taken (the freerun mode). Swift mode also allows for triggered buffered measurements,
in which a host or external trigger indicates when to take each measurement.
The swift mode cannot be entered if a modulated measurement (MAP, PAP, or BAP) is being
performed.
Commands related to the swift mode are based on the SWIFT function code:
Syntax
:
SWIFT [FREERUN or OFF]
[FREERUN] initiates the freerun mode (continuous taking and returning of measure-
ments).
[OFF] causes the 8650A to exit the swift mode; all unread data is lost.
Examples
:
OUTPUT 713;SWIFT FREERUN
! Initiate swift freerun mode
OUTPUT 713;SWIFT OFF
! Exit the swift mode
For commands which set up triggered measurements, the command format is:
Syntax
:
SWIFT [GET or TTL] BUFFER [b]
[GET or TTL] define the trigger:
[GET] the expected trigger is a GPIB GET command. The 8650A signals the host by
asserting SRQ every time it is ready to take a measurement.
[TTL] the expected trigger is a TTL high at the rear panel trigger input.
For triggered measurements, the 8650A signals the host every time it is ready to take a
measurement (consisting of a set of samples equal to the averaging number). The 8650A
then waits for the trigger; when the trigger is received, the 8650A de-asserts the signal to
the host, measures the data, and buffers the data. When the instrument is again ready to
measure data, it again asserts the signal to the host. After the specified number of
measurements, the 8650A is ready to output data.
BUFFER (followed by the numeric variable [b]) specifies the number of measurements to
be taken and stored in the buffer. The minimum value is one. The maximum value is 5000.
Examples
:
OUTPUT 713;SWIFT PRE GET BUFFER 200
! take measurements until GET is received
! then output the last 200 measurements
! taken
OUTPUT 713;SWIFT POST TTL BUFFER 100
! wait for a TTL trigger, then take 100
! measurements
Summary of Contents for 8650A Series
Page 4: ......
Page 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Page 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Page 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Page 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Page 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Page 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Page 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Page 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Page 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...