In BAP mode, the 8650A automatically determines which portions of the signal are in the pulse and
which are not. In computing the average power, the 8650A uses only those portions that are within the
pulse. The result is that, independent of the signal’s pulse duty cycle, the meter always reads the average
power in the pulse or burst. As with the PAP mode, when measuring a pulse modulated signal with 50%
duty cycle, the reading in the BAP mode would be 3 dB higher than in the MAP mode. However, in the
BAP mode, the signal’s duty cycle can change dynamically in time without affecting the meter reading.
In the PAP mode, the duty cycle factor must be entered to match the duty cycle of the pulsed signal.
NOTE:
BAP Mode requires a minimum pulse on or off time as determined by the power
sensor pulse width specification.
☛
☛
☛
☛
Summary of Contents for 8650A Series
Page 4: ......
Page 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Page 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Page 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Page 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Page 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Page 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Page 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Page 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Page 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...