6.8.2.18 Parameter "Flow preset counter
“
Preset counter.
Setting values:
Output signal static
The output changes its state when the preset counter
is reached and remains there until the preset counter
is reset
Output pulse
The output signal changes its state when the prese-
lection value is reached and falls back after an ad-
justable time
6.8.2.19 "Counter duration" parameter
Signal duration. Only visible when Output pulse is selected in the "Preset counter"
parameter.
Setting values: 0.1 ... 100.0 sec.
6.8.2.20 Parameter "Pulse polarity
“
Signal polarity.
Setting values:
Positive
Signal has High level, idle state is low
Negative
Signal has Low level, idle state is high
NOTE
The following parameters are only visible if function Reset input has been selected!
This option is only available for pin 5.