OMNIPLUS-F
Operation | 6
C-E01_X__OMNIPLUS-F__DB2-1.0
27/ 45
6.5.9 Parameter "Start of range | End of range |
Display value"
Only visible and configurable if "Medium" was previously set to "User defined
Is set.
These three parameters are used to adapt the instrument to liquids other than water.
They are only visible if "Medium" has been set to "user defined". The factory adjust-
ment then has no function. With the help of the three parameters, a new characteristic
curve can be defined.
For the procedure for setting the parameters, see
chapter 6.2.2.6 Medium setting.
Setting values:
Start of range
0 ... 1.0000
End of range
0 ... 1.0000
Display value
0 ... 100 %
6.5.10 "Response time" parameter
"Response time" defines the response time of the flow measurement. This is the time
that elapses after a sudden change in the flow velocity until the display has ap-
proached the new value.
Setting values: 0.1 ... 99.9 sec
6.6 "Volume totalizer" submenu
The OMNIPLUS-F has a volume totalizer that can add up the flowed volume depend-
ing on the flow velocity and the set pipe diameter. The flow measurement does not
have to be displayed in a flow unit for this purpose. However, the correct setting of the
pipe diameter is a prerequisite for a meaningful summation.
If the volume counter is activated, it can be displayed in the measured value display of
the main display by
RING-TURN
together with the instantaneous value.
At the same time, a switching signal can be output on pin 4. For this purpose, pin 4
must be configured accordingly (see "Pin 4 settings").
6.6.1 "Totalizer" parameter
Enable or disable the volume totalizer.
Setting values:
ON - activates the volume totalizer display
OFF - turns off the volume totalizer (
the parameters described below will be not
displayed)