![GHM HONSBERG OMNIPLUS-F Series Operating Manual Download Page 12](http://html.mh-extra.com/html/ghm/honsberg-omniplus-f-series/honsberg-omniplus-f-series_operating-manual_2220447012.webp)
5 | Mounting
OMNIPLUS-F
12/ 45
C-E01_X__OMNIPLUS-F__DB2-1.0
5.2 Threaded probe
-
OMNIPLUS-F-008
-
OMNIPLUS-F-015
Threaded sensors have an external thread as process connection, with which they are
screwed into a T-piece or a welding socket of the pipeline. The sealing is done e.g.
with PTFE tape or with liquid seal. Sealing with the aid of a flat gasket is generally not
possible, as the correct alignment of the X-marking is not guaranteed when screwing
in as far as it will go.
Due to the relatively short lengths of the available threaded probes, the range of appli-
cation is limited to nominal diameters from about DN15 to DN50.
The following variants are available:
- Male thread G1/2", length 29.6 mm (type OMNIPLUS-F-015HK029)
- Male thread G1/2", length 45 mm (type OMNIPLUS-F-015HK045)
- External thread G1/4", length 28 mm (type OMNIPLUS-F-008HK028)