6 | Operation
OMNIPLUS-F
30/ 45
C-E01_X__OMNIPLUS-F__DB2-1.0
6.8.1.3
“Analog out min / Analog out max" parameter
These two parameters define the range of the measured value to correspond to the
output range of the analog output. Setting range, resolution and unit correspond to
display range and settings of the assigned measurand (flow or temperature).
See also parameters "Max velocity", "Unit" (configuration level "Flow measuring"),
"Decimals", "Max flow rate" and "Unit" (configuration level Temp measuring).
Parameter "Analog out min
”
Assignment of the measured value MIN for flow or temperature to the analog output
Parameter "Analog out max
”
Assignment of the measured value MAX for flow or temperature to the analog output
6.8.2 "Pin 4 settings" and "Pin 5 settings" submenus
6.8.2.1 "Function" parameter
Setting values:
OFF
Turns off pin 4 or pin 5
Flow switch output
The flow switch allows monitoring of flow velocities or flow
rates for exceeding or falling below adjustable limit values
Temp switch output
The temperature switch allows monitoring of the medium
temperature for exceeding or falling below adjustable limit
values
Flow freq output
The frequency output for flow is an output of the flow ve-
locity or flow rate with an adjustable frequency
Temp freq output
The frequency output for temperature is an output of the
medium temperature with an adjustable frequency
Only for pin 4
Flow pulse output
The pulse output is an output of one pulse each after
passing an adjustable volume
Only for pin 4
Flow preset counter
The output for preselection counter is an output of a signal
after reaching the preselected volume counter reading
Only for pin 5
Totalizer reset input
Reset input for volume counter
Only for pin 5
Inverted pin 4
Inverse signal from pin 4. If pin 4 shows low signal, pin 5
shows high signal and vice versa. Can be used for line
break detection, because pin 4 and 5 must always show
different signals.
Note:
For operation with IO-Link, observe chapter 6.12.