2581 FlowtraMag

®

 Meter 

LED Status Indicators

10 s 

2 s 

20 s 

30 s 

40 s 

50 s 

60 s 

70 s 

NOTE:

 The SENSITIVITY function is ine

 ective if the AVERAGING function is set to o

  (seconds).

No AVERAGING, no SENSITIVITY 

AVERAGING only 

AVERAGING and SENSITIVITY 

With AVERAGING set to O

  and with SENSITIVITY set to zero, the 

2581 FlowtraMag responds to every unstable shift in the 

fl

 ow.  

The dashed red line represents the actual output of the 

fl

 ow sensor 

in unstable 

fl

 ow conditions.

With AVERAGING set to Medium and SENSITIVITY still set to zero 
the 

fl

 ow rate is stabilized, but a sharp change in 

fl

 ow rate is not 

represented for 50 seconds or longer (dotted green line).

With AVERAGING at Medium and SENSITIVITY set to a moderate 
amount, the 

fl

 ow rate is stabilized, while the sudden shift in 

fl

 ow is 

re

fl

 ected very quickly (solid blue line).

• Because ideal 

fl

 ow conditions are often impossible to achieve, the 

fl

 uids 

fl

 ow is often erratic, which causes erratic readings in 

control features (e.g., relays, 4 to 20 mA loops, etc.) that are associated with the 

fl

 ow rate.

•  The best solution to these problems is to correct any piping de

fi

 ciency that causes the instability.  This may involve longer 

straight runs upstream, taking steps to ensure pipe remains full during 

fl

 ow conditions, and other installation changes. 

In many situations, however, these measures are simply not possible.

•  The 2581 FlowtraMag provides two tools that are designed to "work around" these de

fi

 ciencies.  The Averaging and the 

Sensitivity features should be studied before making adjustments.

Averaging Time in Seconds (Factory set: Low)

•  Set the time the meter will use as the averaging period.  The ranges are O

 , Low (10 s), Med (40 s) and High (120 s).

  Use higher averaging times to smooth the display and current output where the 

fl

 ow in the pipe is erratic.

Quick Response Sensitivity (Factory Set:  0.5 m/s (1.64 ft/s) equivalent)

•  

Sets an amount of 

fl

 ow rate change at a given reference 

fl

 ow rate required to momentarily allow the 2581 FlowtraMag to change 

from the selected averaging time to a faster response.  The reference 

fl

 ow rate should be near the range of normal operation. 

Turn the averaging setting to o

  and observe the 

fl

 ow rate variation, enter a sensitivity amount that is two times larger than the 

amount of 

fl

 ow variation. 

Averaging and Sensitivity Settings

LEDs on the 2581 FlowtraMag circuit board are useful to identify problems with the meter and the 

fl

 ow conditions.

LED Condition

Indication

All O

 

The power is o

  or the sensor is not connected

Solid Blue

Normal operation, full pipe, no 

fl

 ow

Blinking Blue

Normal operation, blink rate is proportional to 

fl

 ow rate

Solid Purple

Partially 

fi

 lled pipe, 

fl

 ow rate is zero

Blinking Purple

Partially 

fi

 lled pipe, blink rate is proportional to 

fl

 ow rate

Blinking Red

Measurement out of range. If condition persists, will turn to solid red after 1 minute

Solid Red

Instrument error, defective electronic component. Contact Technical Support

     Green

       - Connected device

     White

       - No connections

Summary of Contents for FlowtraMag GF 2581

Page 1: ...f reading accuracy Partially filled pipe detection status indicator Visual LED indicators make sensor status clear and easy to read Reverse flow direction configurable with 0252 Configuration Tool or...

Page 2: ...1 Depressurize and vent system prior to installation or removal 2 Confirm chemical compatibility before use 3 DO NOT exceed maximum temperature or pressure specs 4 ALWAYS wear safety goggles or face...

Page 3: ...ator Digital S3 L Output Digital S3 L 4 5 to 5 5 VDC Serial ASCII TTL level 9600 bps The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by...

Page 4: ...n 1 41 in 2X 0 22 in 3 91 in 5 46 in 9 82 in 7 86 0 062 in 199 6 1 6 mm 213 7 1 8 mm Pipe Cutout dimension 2X 28 8 mm 25 5 mm 35 8 mm 2X 5 5 mm 99 2 mm 249 4 mm 138 7 mm 3 91 in 9 82 in 5 46 in 2X 0 2...

Page 5: ...G ASTM and Metric pipe cutout dimensions are di erent DN80 Flange bolt kits and gaskets not shown Sold separately DN100 Flange bolt kits and gaskets not shown Sold separately 3 91 in 10 90 0 125 in 9...

Page 6: ...the 12 o clock position OK Recommended OK Vertical flow is OK IF the pipe remains FULL at all times DO NOT HANDLE BY THE SENSOR HEAD Always handle FlowtraMag Meters by the union nuts or flanges NOT t...

Page 7: ...nts 2 Select appropriate Metric or ASTM union end for installation 3 Install sensor with flow arrow pointing in the direction of flow Note Gland fittings should point upstream of flow 2581 FlowtraMag...

Page 8: ...440 713 FKM Profile Flange Gasket Metric 2 DN100 Metric Bolt Pattern Components Note Use M16 bolts Sold Separately Part Number Description Qty Req d 721 790 114 DN100 Flange Adapter PVC U Metric Use b...

Page 9: ...ronments connect the power cable shield drain wire to a clean low impedance earth ground If there is a single power supply for all 2581 FlowtraMag system components Instruments PLCs and VFDs route sig...

Page 10: ...g Tool App Default Configurations Wiring Configuration Continued DN25 1 in DN40 1 5 in DN50 2 in DN80 3 in DN100 4 in Units GPM GPM GPM GPM GPM Totalizer Units Gallons Gallons Gallons Gallons Gallons...

Page 11: ...ERED Signet Flow Instruments CAUTION Turn o Power before Wiring DO NOT REMOVE Seal Warranty void if seal is broken or removed WARRANTY VOID IF REMOVED Black Red Shield EMC enhancement 24 VDC 1A min Po...

Page 12: ...OR S3L Input 2 S3L Input 1 Frequency 1 S3 L 2 Ground White Data Red Output cable Black Brown Shield Blue not used Not used OR OR 2581 Ground White Data Red Output cable Black Brown Shield Blue not us...

Page 13: ...ual for Frequency wiring and programing instructions Wiring with Digital S3 L Output PWR 9900 Power Input LOOP LOOP PWR 9900 S3 L Terminal DATA GND V SHLD 9900 Transmitter S3L Freq 24 VDC 1A min Power...

Page 14: ...rn to home screen and look for App icon click the blue GF Config Tool icon 3 Continue to Sensor Setup Section next page Bluetooth App Setup Steps Android version Download the GF Configuration Tool App...

Page 15: ...ear Settings example shows Menu Android version Version 8 or newer When the 2581 FlowtraMag is in operation when in close proximity to the 2581 FlowtraMag less than 20 ft open the GF Config Tool App t...

Page 16: ...ly logs current screen view in real time when connected to the app App Configuration Monitor and Real Time Log Monitoring flow and totalizer Live Logging while connected to mobile tablet device set 1...

Page 17: ...saved settings configuration file See Save function above NOTE The file must be a GF Config Tool App settings configuration file The software will verify whether the user selected file is the correct...

Page 18: ...tal S3 L and Frequency and or 4 to 20 Active or Passive use the GF Config Tool App On the loop screen use the drop down to select digital S3 L or Freq and or Active or Passive 4 to 20 mA Press WRITE a...

Page 19: ...cable Black Brown Shield Blue not used Not used Ground White Data Red Bluetooth set to S L iOS device with home button Double click the home button Find the minimized app and swipe up to close the app...

Page 20: ...o tools that are designed to work around these deficiencies The Averaging and the Sensitivity features should be studied before making adjustments Averaging Time in Seconds Factory set Low Set the tim...

Page 21: ...tivity Low Flow Cut O and Direction of the Flow drop down menu 4 If the user has any correction or change you must Write the information to device then click Read device again 1 2 This is an outline F...

Page 22: ...on or change you must Write the information to device then click Read device again 5 The Loop tab select or confirm your 4 to 20 mA set point set your current alarm condition and type of output mode I...

Page 23: ...Tool 7 8 8 The Calibration tab allows custom calibration via method of rate volumetric zero flow calibration and reset flow calibration 7 The Monitor tab can graph or log the information to your local...

Page 24: ...0252 Configuration Tool 2 Check wiring Use the wiring diagram picture in the 2581 FlowtraMag manual Frequency Digital or Current Output is Erratic 1 Electrical Noise interference with the flow measure...

Page 25: ...th units present Record the FlowtraMag s IDs or delete unused sensors 3 Restart GF Config Tool App 4 to 20 mA output is incorrect 1 The 4 to 20 mA in the 2581 FlowtraMag is not scaled properly The Ins...

Page 26: ...Notes...

Page 27: ...aMag PVC Hastelloy C FKM O Ring Flange DN80 3 in 3 2581 PH04 101 159 001 979 FlowtraMag PVC Hastelloy C FKM O Ring Flange DN100 4 in 3 2581 PT01 102 159 001 980 FlowtraMag PVC Titanium EPDM O Ring Uni...

Page 28: ...2581 PX02 10X Accessories 857 375 020 857 375 020 PVC 80 Type 375 Union FKM SxS 2 in ASTM 1220 0331 159 812 041 2 225X 210 FKM O RING RMS1071 2 in 2 required per unit 897 375 020 897 375 020 PVC 80 T...

Reviews: