![Gentherm CSZ Blanketrol II 222S Operation And Technical Manual Download Page 81](http://html1.mh-extra.com/html/gentherm/csz-blanketrol-ii-222s/csz-blanketrol-ii-222s_operation-and-technical-manual_2213792081.webp)
FIELD REPAIR/SERVICE
BLANKETROL II, Model 222S
OPERATION AND TECHNICAL MANUAL
Page 81 of 110
H. Remove the four remaining Phillips head screws around the face of the water flow
indicator. Set the screws to the side.
I. Disassemble the parts of the water flow indicator. Do not lose the large black O-ring.
J. Rinse thoroughly with clean water. A small brush may be used to clean the parts.
K. Reassemble the parts.
L. Reinsert the four screws.
M. Position the replacement water flow indicator inside the unit so that the screw holes are
aligned on the right side wall and the plastic elbows point toward the back of the unit.
N. Insert and tighten the two Phillips head screws from the outside.
O. Connect the hose from the upper manifold to the top left plastic elbow and tighten the
clamp.
P. Connect the hose from the water filter to the top right plastic elbow and tighten the
clamp.
Q. Refill the water reservoir as described in Section (4-2.5.).
R.
Reposition the front storage drawer as described in Section (5-1.6).
5-12. REPLACEMENT OF THE I/O POWER SWITCH
A. Obtain the replacement I/O power switch.
B. Extend the front storage drawer as described in Section (5-1.5.).
C. Locate the square I/O power switch on the left side of the front panel of the unit.
Number 13 in Figure (6-2.).
D. Work the power switch and attached cable to the outside of the unit by pressing together
the tension clips on the top and bottom of the switch assembly. The tension clips are
accessible by reaching inside the unit from the extended drawer opening.
E. Examine the placement of the wires on the power switch.
F. Disconnect the four slide-on connectors from the switch.
G. Position the replacement power switch into the unit so that the labels are upright and
“I”
is at the top.
WARNING
Always
unplug the unit
before accessing internal components during service.
Failure to unplug the unit could result in electric shock.
Summary of Contents for CSZ Blanketrol II 222S
Page 112: ......