After arranging the equipment as described in Section (3-1.), proceed as follows:
A. Press the power switch to
the “I” position.
1. The switch lights up.
2. The microprocessor board goes through self-test.
3. The SETPOINT/ STATUS display flashes CHECK SETPT.
B. Consult the physician's orders to determine the desired patient temperature and the
desired water temperature setting.
C. Press the TEMP SET Switch.
1. The microprocessor board beeps.
2. The LED in the corner of the button lights up.
3. The SET POINT display shows a temperature reading.
4. The SETPOINT/ STATUS display shows SET TEMP.
D. Use the C/F button to choose the desired temperature scale. (English Membrane Only)
E. Press the Up arrow or Down arrow to change the SETPOINT display to the desired
Water set point temperature. As a safety precaution, the Water temperature in the
device can only be set between 4°C - 42°C (39.2°F - 107.6°F).
1. The microprocessor board beeps.
2. The SETPOINT/ STATUS display changes.
F. Press the MANUAL CONTROL switch.
1. The microprocessor board beeps.
2. The LED in the corner of the button lights up.
3. The WATER display shows the actual temperature of the water in the
BLANKETROL II equipment.
4. The SETPOINT/ STATUS display shows:
XXXXXX WATER
MANUAL SETPT 37.0 C
Or SETPOINT/ STATUS display shows:
WATER @SETPT
MANUAL SETPT 37.0 C
NOTE
: “XXXXXX
” is HEATING, or COOLING. The above depends upon the
relationship of the blanket water temperature to the set point temperature.
NOTE (ENGLISH MEMBRANE ONLY):
SET POINT may be displayed in
Fahrenheit if the operator is in Fahrenheit mode.
5. The pump is activated. The heater or compressor may be activated.
6. The Water Flow indicator on the right side panel begins to move.
7. The water moves from the unit through the blanket and returns to the unit.
G. Check the water flow indicator to confirm that the water is circulating.
H. Touch the hyper-hypothermia blanket to confirm that the blanket is heating/cooling.
I. To make any changes in the control setting, press the TEMP SET button and begin
again.