After arranging the equipment as described in Section (3-1.), proceed as follows:
A. Check the placement of the 400 Series probe in or on the patient.
B. Insert the 400 Series probe into the probe jack on the right side of the unit.
C. Press the power switch to
the “I” position.
1. The switch lights up.
2. The microprocessor board goes through self-test.
3. The SETPOINT/ STATUS display flashes CHECK SETPT.
D. Consult the physician's orders to determine the desired patient temperature. As a safety
precaution, the SET POINT display can only be set between 30°C - 40°C (86°F - 104°F)
to operate in
AUTO CONTROL MODE
.
E. Use the C/F button to choose the desired temperature scale. (English Membrane only)
F. Press the TEMP SET button.
1. The microprocessor board beeps.
2. The LED in the corner of the button lights up.
3. The SET POINT display shows a temperature reading.
4. The SETPOINT/ STATUS display shows SET TEMP.
G. Press the Up arrow or Down arrow to change the SET POINT display to the desired
patient temperature. The display can only be set between 30°C -40°C (86°F - 104°F) or
else the unit will not operate in
AUTO CONTROL MODE.
1. The microprocessor board beeps.
2. The set point on the SETPOINT/ STATUS display changes.
H. Press the AUTO CONTROL button.
1. The microprocessor board beeps.
2. The LED in the corner of the button lights up.
3. The Patient display shows the patient's actual temperature.
4. The WATER display shows the actual temperature of the water in the
BLANKETROL II equipment.
5. The SETPOINT/ STATUS display indicates:
XXXXXX PATIENT
AUTO SETPT 37.0 C
Or SETPOINT/ STATUS display shows:
PATIENT @SETPT
AUTO SETPT 37.0 C
NOTE
: “XXXXXX
” is HEATING, or COOLING. The above depends upon the
relationship of the patient temperature to the set point temperature.
NOTE (ENGLISH MEMBRANE ONLY):
SET POINT may be displayed in
Fahrenheit if the operator is in Fahrenheit mode.
6. The pump is activated. The heater or compressor may also be activated.
7. The water flow indicator on the right side panel begins to move.
8. The water moves from the unit to the blanket and returns to the unit.