![Gentherm CSZ Blanketrol II 222S Operation And Technical Manual Download Page 48](http://html1.mh-extra.com/html/gentherm/csz-blanketrol-ii-222s/csz-blanketrol-ii-222s_operation-and-technical-manual_2213792048.webp)
OPERATION
BLANKETROL II, Model 222S
OPERATION AND TECHNICAL MANUAL
Page 48 of 110
HIGH LIMIT
REMOVE FROM SERVICE
This message occurs when the software fails and the water in the
BLANKETROL II equipment has reached the high temperature limit of
44.0°C
2°C (111.2°F
3.6°F), activating the independent safety. While
this message is displayed, the trouble alarm will sound, and the heater
and pump will turn off. The alarm silence key cannot silence this alarm.
This message will continue to display and the trouble alarm will continue
to sound until the unit is turned off using the ON/OFF switch or the power
cord is disconnected from the power source. If the unit is unplugged
before turning the ON/OFF switch, the power fail alarm will activate and
the power fail LED will light. If the unit is turned back on, this message
will reappear and the trouble alarm will sound. The unit cannot be used
again until it is serviced. The seven-seg
ment displays (“WATER” and
“PATIENT”) will both be blank during this condition. Contact CSZ
Technical Service.
SNAP DISC
REMOVE FROM SERVICE
This message occurs when the thermostatic snap disc has opened or is
disconnected, or when the reservoir temperature has exceeded 46°C
2°C
(114.8
3.6°F) causing the thermostatic disc to shut off heating. While this
message is displayed, EE02** will display in the “PATIENT” temperature
display, the trouble alarm will sound, and the heater, compressor and
pump will turn off. The alarm silence key cannot silence this alarm. This
message will continue to display and the trouble alarm will continue to
sound until the unit is turned off using the ON/OFF switch or the power
cord is disconnected from the power source. If the unit is unplugged
before turning the ON/OFF switch, the power fail alarm will activate and
the power fail LED will light. If the unit is turned back on,
only
”SNAP
DISC” “REMOVE FROM SERVICE” will reappear and the trouble alarm
will sound, even if disc has reset.
The unit cannot be used again until it is
serviced. Contact CSZ Technical Service.
**EE02 appears when the thermostatic snap disc has opened or is
disconnected.
LOW LIMIT
REMOVE FROM SERVICE
This message occurs when the low limit is triggered because the water
in
the BLANKETROL II equipment has reached the low temperature limit of
2.0°C
2°C (35.6°F
3.6°F). While this message is displayed, the
trouble alarm will sound, and the compressor and pump will turn off. The
alarm silence key cannot silence this alarm. This message will continue to
display and the trouble alarm will continue to sound until the unit is turned
off using the ON/OFF switch or the power cord is disconnected from the
power source. If the unit is unplugged before turning the ON/OFF switch,
the power fail alarm will activate and the power fail LED will light. If the
unit is turned back on, this message will reappear and the trouble alarm
will sound. The unit cannot be used again until it is serviced. The seven-
segment displays (“WATER” and ”PATIENT”) will both be blank during
this condition. Contact CSZ Technical Service.
LOW LIMIT
REMOVE FROM SERVICE
This message occurs when the software fails and the water in the
BLANKETROL II equipment has reached the low temperature limit of
2.0°C
2°C (35.6°F
3.6°F), activating the independent safety. While
this message is displayed, the trouble alarm will sound, and the
compressor and pump will turn off. The alarm silence key cannot silence
this alarm. This message will continue to display and the trouble alarm
will continue to sound until the unit is turned off using the ON/OFF switch
or the power cord is disconnected from the power source. If the unit is
unplugged before turning the ON/OFF switch, the power fail alarm will
activate and the power fail LED will light. If the unit is turned back on, this
message will reappear and the trouble alarm will sound. The unit cannot
be used again until it is serviced. The seven-
segment displays (“WATER”
and
“PATIENT”) will both be blank during this condition. Contact CSZ
Technical Service.
Summary of Contents for CSZ Blanketrol II 222S
Page 112: ......