29
TS4000
NOTE:
If the TS4000 is ordered without relays or MODBUS communications, changing the
relay or MODBUS settings will have no effect on the operation of the unit.
4.3 User Menu Display
The following table explains the User Menu abbreviations displayed on the 3-digit 7-segment
LED display:
Table 15: User Menu Display
User
Menu
Display
Definition
Top
Menu
rSt
Reset Relays
---
Gas Check
AC
Calibration
SE
Setup
Fi
Finish, exiting at any level of the menu to the upper level
r
Return from top menu to normal operation
Reset Sub Menu
C
Clearing relays, appears for 2 seconds
Gas Check Sub Menu
---
Flashing, unit is zeroing in Gas Check
gAS
Flashing, unit is finished zeroing and ready for gas to be applied
###
Flashing gas reading, unit is reading gas in Gas Check
Calibration Sub Menu
###
Flashing remaining sensor life – Unit is zeroing in Calibration
100
Remaining sensor life was reset to 100, unit is still zeroing
gAS
Flashing – Unit is done zeroing and is ready for gas to be applied
CP
Flashing – Unit is seeing gas, calibration is in progress
CC
Steady – Unit has finished calibration and telling user to remove the gas
Setup Menu
Sr
Sensor scale / type setting
O2
Oxygen deficiency sensor
CO
Carbon Monoxide sensor (range settings: 100 ppm and 500 ppm)
100
100 ppm scale
500
500 ppm scale
NH3
Ammonia sensor (range settings: 50 ppm and 100 ppm)
50
50 ppm scale
100
100 ppm scale
Cl2
Chlorine sensor (range settings: 10 ppm and 20 ppm)
10
10 ppm scale
20
20 ppm scale
LO
Warning relay setting (Alarm setting for O
2
)
dE
De-energize relay
En
Energize relay
nL
Non-latching relay
LA
Latching relay