9
Cleaning
9.1
Cleaning
The valve is suitable for CIP (Cleaning in Place); recommended cleaning speed
in the valve is at least 2 m/s.
All parts in contact with product must be cleaned at regular intervals. Always
observe the safety data sheets issued by the cleaning agent manufacturers. Only
use cleaning agents which do not cause damage to the seals and the inner parts
of the valve.
With respect to the cleaning method and parameters like detergents,
temperatures, times, and intervals, the component manufacturer can merely
make recommendations but cannot provide any generally applicable details.
Method and parameters should be determined and defined by the operator in
accordance with the relevant process and product.
The cleaning effect must be checked regularly by the operating company!
9.2
Sterilisation
The valve is suitable for SIP sterilisation (Sterilisation in Place). For valves
equipped with the sealing material EPDM, the following specifications apply.
Sterilisation is possible with:
•
Hot water at max. 150 °C (302 °F) for 20 to 30 min
•
Steam at max. 150 °C (302 °F) for 20 to 30 min
•
Chemicals (e.g. H
2
O
2
)
9.3
Passivation
Before commissioning a plant, passivation is commonly carried out for long pipes
and tanks. Valve blocks are usually excepted from this.
Passivation is typically performed using nitric acid (HNO
3
) at approx. 80 °C (176
°F) at a concentration of 3 % and a contact time of 6 to 8 hours.
The ultimate temperatures, chemicals, concentrations and contact time to be
used must be determined by the plant operator along with its chemical supplier.
Cleaning
Cleaning
430BAL013317EN_1
11.03.2019
29
Summary of Contents for Aseptomag AMV
Page 60: ......