Water bath
Carry out the following steps:
1.
Insert the pressurizing tool into the inner assembly.
Fig.26:
Inserting the pressurizing tool
2.
Tightly screw the pressurizing tool to the inner assembly by hand.
!
With the inner assembly, ideally apply the force to the valve seat.
Fig.27:
Screwing the pressurizing tool
3.
Apply max. 3 bar of compressed air to the pressurizing tool.
!
Air pressures > 3 bar can damage the metal bellows.
4.
Immerse the internal assembly in a water bath for approx. 30 seconds.
!
While the internal assembly is immersed, check the internal assembly for
leaks. Leakages will be indicated by air bubbles appearing on the surface.
Maintenance
Carrying out the "Internal assembly" leak test (bubble test)