DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
10-8
Chapter 10 - Care and Maintenance
P R E L I M I N A R Y
10-5-2-2
Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted.
Refer to the
Venue Fit™ User Manual
for a list of approved peripherals/options.
10-5-2-3
Mains Cable Inspection
Table 10-6 GE Approved Peripheral/Hardware Option Functional Checks
Step
Item
Description
1
Media
Verify media drive(s) read/write properly.
2
B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
Table 10-7 Mains Cable Inspection, As Appropriate
Step
Item
Description
1
Unplug Cord
Disconnect the mains cable from the wall outlet and from the Ultrasound system.
2
Mains Cable
Inspect the mains cable and its connectors for any damage.
3
Terminals
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals,
and that no strands may cause a short circuit.
4
Inlet Connector
Verify that the Inlet connector retainer is functional.
Summary of Contents for Venue Fit