Chapter 2 - Site Preparations
P R E L I M I N A R Y
2-2-5-5
Power Stability Requirements
•
Power Transients
The Venue Fit™ is fully compliant with the following standard:
IEC61000-3-3 and EN\IEC60601-1-2.
2-2-6
EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also generate EMI. The Venue Fit™ ultrasound
unit complies with limits as stated on the EMC label. However, there is no guarantee that interference
will not occur in a particular setup.
NOTE:
Possible EMI sources should be identified before the unit is installed, and should not be on the same
line as the ultrasound system. A dedicated line should be used for the ultrasound system.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Some of
these sources include:
•
medical lasers
•
scanners
•
cauterizing guns
•
computers
•
monitors
•
fans
•
gel warmers
•
microwave ovens
•
light dimmers
•
mobile phones
•
in-house wireless phones (DECT phones)
•
wireless computer keyboard and mouse
•
air conditioning system
•
High Frequency (HF) surgery equipment
•
general AC/DC adapters
The presence of a broadcast station or broadcast van may also cause interference.