DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
4-12
Chapter 4 - General Procedures and Functional Checks
P R E L I M I N A R Y
4-3-3-1-5
12L-RS (Linear) Probe Test
16) Connect 12L-RS probe
17) Select Artery preset
18) Run scanning on air scan in B-mode
19.)Increase depth to max
20) Increase gain till image is visible
21.)Verify image is homogeneous, no artifacts or structured noise appears.
Figure 4-47 2D B-Mode Image Test - 12L-RS Probe