![GE ReadyToProcess WAVE 25 Operating Instructions Manual Download Page 161](http://html.mh-extra.com/html/ge/readytoprocess-wave-25/readytoprocess-wave-25_operating-instructions-manual_128043161.webp)
7.3
ReadyToProcess CBCU
For more detailed troubleshooting, see the ReadyToProcess WAVE 25 System Handbook.
Corrective action
Possible cause
Error symptom
Check any warning message and follow the instruc-
tion. If problem persists, contact GE service person-
nel.
An internal error has
occurred, but the CBCU
is still operating.
Status LED flashes red.
Check any warning message and follow the instruc-
tion. If problem persists, contact GE service person-
nel.
An internal error has
occurred, and the CBCU
is not operating.
Status LED shows a
steady red light.
Contact GE service personnel.
An internal error has
occurred, and the CBCU
is not operating proper-
ly.
Note:
This is normal the first
seconds during power
up.
CAN indicator LED
flashes.
ReadyToProcess WAVE 25 Operating Instructions 29009597 AD
161
7 Troubleshooting
7.3 ReadyToProcess CBCU