April 2004
3-4
Advanced Programming
Setting Up the Alarms
The Series 1 has two optional alarms for each channel. To select the
measurement mode, units and set point values for each alarm, press
the keys shown in Table 3-2. A screen appears similar to Figure 3-2
below. Figure 3-19 on page 3-32 for a menu map.
To test alarms, refer to Testing the Alarm Relays in Chapter 2,
Troubleshooting and Maintenance, of the Service Manual.
*The Settings Menu will prompt for a passcode (see page 2-6).
Figure 3-2: Alarm Settings Menu
Note:
The first time you enter this menu, all alarm relays are OFF.
When an alarm value is not entered, the alarm relay is OFF.
Table 3-2: Moving to Alarm Settings Menu
Press the Main Menu key:
To enter the:
SETTINGS*
Settings Menu
OUTPUTS
Output Settings Menu
ALARMS
Alarm Settings Menu
SET CHANNEL TEST DONE
Alarm Settings Menu
ALARM SETPOINT
Ch
Alarm A Alarm B
Mode: OFF OFF
Trip: Above Above
Setpoint:
+0.0000 +0.0000
+0.0000 +0.0000
Deadband:
Status:
Reset
Reset
1