![GE Moisture Image 1 Series Programming Manual Download Page 43](http://html.mh-extra.com/html/ge/moisture-image-1-series/moisture-image-1-series_programming-manual_128402043.webp)
April 2004
2-24
Basic Programming
Entering Delta F Oxygen
Cell Reference Values
IMPORTANT:
The instructions in this section apply to entering
reference data for the Delta F oxygen cell only.
Use the steps below to change the reference values for oxygen cell
inputs. (Figure 3-19 on page 3-32 for a menu map.)
Caution!
Do not adjust oxygen cell reference data unless
instructed to do so by the factory.
IMPORTANT:
You should record this data on the Program
Information List supplied in Appendix A of the
Startup Guide, or on a separate sheet of paper, and
keep it in a safe place.
1. At the Reference Menu (see Table 2-4 on page 2-21), press the
PROBE menu key until the Oxygen Cell Reference Table appears
(see Figure 2-12 below).
Figure 2-12: Reference Menu for the Delta F Cell
2. Compare the data on the screen to the reference data printed on the
label placed on the side or front of the unit. If the data is incorrect,
use the following procedure to correct it. If the data is correct, do
one of the following:
•
continue to enter reference values for moisture and pressure
inputs as described in Entering Moisture Reference Values on
page 2-22, and Entering Pressure Reference Values on
page 2-26.
•
press the DONE menu key until Main Menu appears on the
message line; then refer to Displaying Measurements on
page 2-28. When you exit, the Series 1 automatically calibrates
each channel.
OXYGEN CELL REFERENCE TABLE
HIGH LOW
1
2
3
4
5
6
AutoCal Interval: (HH.MM): 01:00
Reference Menu 11:15 6/2/00
HELP INTERVAL PROBE DONE
Ch Zero Span Zero Span