Basic Programming
2-17
April 2004
Entering Pressure
Calibration Data
Use this section to enter pressure calibration data for any type of
pressure sensor you are using, even if the pressure sensor is connected
to an auxiliary input.
To enter pressure calibration data, you must enter
•
the probe serial number
•
the zero and span range in mV, current, or FP
•
the zero and span range in PSIg
Referring to the Calibration Data Sheets supplied by GE Panametrics
or another manufacturer, enter calibration data separately for each
probe on the designated channel as described below. Figure 3-19 on
page 3-32 for a menu map.
IMPORTANT:
Staple Calibration Data Sheets to the Program
Information List (Appendix A of the Startup Guide)
and store them in a safe place. If they are lost,
contact GE Panametrics for a duplicate. If you alter
any of the calibration data, make sure you make the
change on the Calibration Data Sheet.
Note:
If you want to enter a constant value, refer to Entering
Constants and User Functions on page 3-9. It is not necessary
to enter the calibration data if you are using a constant.
1. At the System Calibration Menu (see Table 2-3 on page 2-12),
press the PROBE menu key until the Pressure Probe Calibration
screen appears. See Figure 2-9 below.
Figure 2-9: System Calibration Menu for Pressure
HELP CHANNEL PROBE DONE
Pressure Probe Calibration Ch
1
S/N
mV Pressure, PSIg
Zero: ____ ____
Span: ____ ____
System Calibration Menu