GE Moisture Image 1 Series Programming Manual Download Page 118

Index

5

April 2004

Index (cont.)

S

Sample Output

Channel  . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Description . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Mode/Units. . . . . . . . . . . . . . . . . . . . . . . . 3-24

Saturation Constants

in PanaView . . . . . . . . . . . . . . . . . . . . . . . 4-15

Saturations . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Screen

Adjusting Contrast . . . . . . . . . . . . . . . . . . 2-40
Displayed Measurement Modes . . . . . . . . 2-31
Displayed Units  . . . . . . . . . . . . . . . . . . . . 2-31
Displaying Measurements  . . . . . . . . . . . . 2-28
Matrix Format, Setting Up . . . . . . .  2-32, 2-34
Scrolling through Pages . . . . . . . . . . . . . . 2-34

Screen Functions  . . . . . . . . . . . . . . . . . . . . . . 2-5
Scrolling through Pages . . . . . . . . . . . . . . . . 2-34
Setting  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
Setting Up. . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22

Activating and Changing Probes . . . . . . . . 2-9
Calibration Data, Entering . . . . . . . . . . . . 2-12
Data Logger . . . . . . . . . . . . . . . . . . . . . . . 3-17
High and Low Reference Values  . . . . . . . 2-21
Matrix Format  . . . . . . . . . . . . . . . .  2-32, 2-34
Verifying. . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8

Setting Up a Box  . . . . . . . . . . . . . . . . . . . . . 2-33
Settings Menu

Alarm Settings Menu . . . . . . . . . . . . . . . . . 3-4
Auto-Cal Interval, Activating. . . . . . . . . . . 3-7
Error Processing Menu. . . . . . . . . . . . . . . 3-28
Passcode . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Port Settings Menu . . . . . . . . . . . . . . . . . . 3-22
Probe Configuration Menu. . . . . . . . . . . . . 2-9
Recorders Settings Menu . . . . . . . . . . . . . . 3-1
Reference Menu . . . . . . . . . . . . . . . . . . . . 2-21
System Calibration Menu. . . . . . . . . . . . . 2-12
System Service Menu. . . . . . . . . . . . . . . . 3-15

Software, Loading  . . . . . . . . . . . . . . . . . . . . 3-30
Specifications

Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8

Specifying User Functions . . . . . . . . . . . . . . 3-11
System Service Menu. . . . . . . . . . . . . . . . . . 3-15

T

Tables

Entering in PanaView . . . . . . . . . . . . . . . .4-14

Tables, User-Defined  . . . . . . . . . . . . . . . . . .3-13
Tagging Inputs  . . . . . . . . . . . . . . . . . . . . . . .2-37
Testing Recorder Outputs

in PanaView  . . . . . . . . . . . . . . . . . . . . . . .4-10

TF Series Probe

Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Description of . . . . . . . . . . . . . . . . . . . 1-4, 1-5

Top Drain Cell

Description of . . . . . . . . . . . . . . . . . . . . . . .1-7

Trimming Recorder Outputs

in PanaView  . . . . . . . . . . . . . . . . . . . . . . .4-10

U

Unassigned Box

How to Make. . . . . . . . . . . . . . . . . . . . . . .2-33

Units

Displayed Units. . . . . . . . . . . . . . . . . . . . .2-31
Line Graph, Selecting . . . . . . . . . . . . . . . .2-37
Matrix Format, Selecting  . . . . . . . . . . . . .2-33
Measurement Modes . . . . . . . . . . . . . . . . .2-31

User Constants

Definition  . . . . . . . . . . . . . . . . . . . . . . . . . .3-9

User Functions

in PanaView  . . . . . . . . . . . . . . . . . . . . . . .4-12

User Functions, Definition of . . . . . . . . . . . . .3-9
User Functions, Entering  . . . . . . . . . . . . . . . 3-11
User Functions, Specifying . . . . . . . . . . . . . . 3-11
User Program

Description of . . . . . . . . . . . . . . . . . . . . . . .1-9
Entering Data  . . . . . . . . . . . . . . . . . . . . . . .2-4

User Tables, Entering  . . . . . . . . . . . . . . . . . .3-13
User-Defined Tables

Entering in PanaView . . . . . . . . . . . . . . . .4-14

V

Verifying Setup Data. . . . . . . . . . . . . . . . . . . .2-8

W

Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .iii

Y

YES Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6

Summary of Contents for Moisture Image 1 Series

Page 1: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Page 2: ...Moisture Image Series 1 Hygrometer Programming Manual ...

Page 3: ...isture Image Series 1 Hygrometer Programming Manual 910 108PB2 ATTENTION This manual contains instructions for Series 1 units equipped with controller card p n 703 1250 This controller card supports the PanaCom PanaView user interface software ...

Page 4: ...d warranties arising from course of dealing or usage or trade Return Policy If a GE Panametrics instrument malfunctions within the warranty period the following procedure must be completed 1 Notify GE Panametrics giving full details of the problem and provide the model number and serial number of the instrument If the nature of the problem indicates the need for factory service GE Panametrics will...

Page 5: ...e Sensors 1 8 RS232 Communications Port 1 8 User Program 1 9 Chapter 2 Basic Programming Introduction 2 1 Getting Started 2 2 Powering Up 2 2 Entering Data into the User Program 2 4 Screen and Key Functions 2 5 Entering the Passcode 2 6 Getting On Line Help 2 7 Verifying and Entering Setup Data 2 8 Activating and Changing Probes 2 9 Entering Calibration Data 2 12 Entering High and Low Reference Va...

Page 6: ...ting Up the Data Logger 3 17 Viewing the Data Logger Status 3 18 Viewing Logged Data 3 18 Removing Logged Data 3 18 Entering Data Logger Settings 3 19 Setting Up the RS232 Communications Port 3 22 Configuring the Serial Port 3 22 Description of a Sample Output 3 24 Setting Error Processing 3 28 Procedure for Setting Error Processing 3 28 Loading New Software 3 30 Chapter 4 Programming with PanaVie...

Page 7: ...Chapter 1 ...

Page 8: ...Features and Capabilities Introduction to the Moisture Image Series 1 1 1 Electronics Unit 1 1 Probes 1 4 Cabling 1 8 User Program 1 9 ...

Page 9: ...s discuss each part of this system Electronics Unit The center of the system is the electronics unit The electronics unit has many components including terminal blocks for all connections an RS232 port a Liquid Crystal Display LCD a keypad and menu keys See Figure 1 1 The back panel of the electronics unit contains the terminal blocks for connecting moisture probes oxygen cells recorders and optio...

Page 10: ...The function of a menu key is indicated directly above it on the LCD The menu key functions change as the user makes menu selections The electronics unit with all these features is available in a rack mount bench mount panel mount weatherproof or explosion proof versions Figure 1 1 The Series 1 Electronics Unit Features P O W E R 1 2 3 Y E S 4 8 6 N O 7 9 5 0 12 250 3 RS232 Port Terminal Blocks LC...

Page 11: ... in IEC CENELEC zone 0 areas CSA Certification Newer versions of rack bench panel and WPF Series 1 s are CSA NRTL certified intrinsically safe Installation requires the Series 1 be mounted in either a NEC CL 1 Div 2 Group B C or D hazardous area or a general purpose area M Series TF Series and Moisture Image Series probes and Delta F oxygen cells are intrinsically safe when connected to the Series...

Page 12: ... in more detail below M Series and TF Series Moisture Probes The M Series and the TF Series probes are very similar Both probes use GE Panametrics aluminum oxide sensors to measure moisture and thermistor beads to measure temperature the thermistor is optional The TF Series however also has an optional built in strain gage type pressure transducer The sensor assemblies are secured to a probe mount...

Page 13: ... see Chapter 3 Specifications of the Startup Guide for BASEEFA certification numbers and designed to meet the requirements of IEC CENELEC zone 0 areas When operated with a CSA NRTL approved Series 1 the M and TF Series probes are intrinsically safe see Chapter 3 Specifications of the Series 1 Startup Guide for CSA NRTL certification numbers and designed to meet the requirements of NEC C1 1 Div 1 G...

Page 14: ...lly safe see Chapter 3 Specifications of the Startup Guide for BASEEFA certification numbers and designed to meet the requirements of IEC CENELEC zone 0 areas The Moisture Image Series Probe when operated with a CSA NRTL certified Series 1 is intrinsically safe see Chapter 3 Specifications of the Series 1 Startup Guide for CSA NRTL certification numbers and designed to meet the requirements of NEC...

Page 15: ... Delta F Cell is available in a general purpose model with both a top and bottom drain or only a top drain see Figure 1 5 below The Delta F Cell is usually installed in its own sample system The Delta F Cell is available with VCR fittings mounted in a NEMA 4X enclosure for weatherproof corrosion resistant applications or mounted in a NEMA 7 enclosure for hazardous areas The Delta F Cell when opera...

Page 16: ...ng 22 AWG twisted pair cable Consult GE Panametrics for longer cable lengths Delta F Oxygen Cells The standard Delta F Cell is connected to the analyzer with a four wire shielded cable 22 AWG With 22 AWG cable oxygen cells with a range from 0 to 1 10 100 ppmv or 0 to 0 5 5 50 ppmv can be located up to 90 meters 300 feet away from the analyzer All other oxygen cells can be located up to 15 meters 5...

Page 17: ...m consists of the following four main menus HELP provides on line help for the various menus within the user program SELECT lets the user select the type of measurement and units to display SETTINGS enables the user to enter probe data and set up basic functions such as recorders and alarms OPTIONS enables the user to set up the Series 1 screen to display measurements in matrix or graphic format a...

Page 18: ...Chapter 2 ...

Page 19: ...Basic Programming Introduction 2 1 Getting Started 2 2 Verifying and Entering Setup Data 2 8 Displaying Measurements 2 28 Adjusting the Contrast of the Screen 2 40 Setting the Clock and Calendar 2 41 ...

Page 20: ...s cover basic programming functions Getting Started describes how to power up gives a description of the user program and explains how to obtain on line help Verifying Setup Data describes how to verify and if necessary enter setup data for moisture pressure oxygen and any other input devices Displaying Measurements explains how to display measurements using different screen formats Adjusting the ...

Page 21: ...RIBED IN THIS MANUAL IN ADDITION BE SURE TO FOLLOW ALL APPLICABLE SAFETY CODES AND REGULATIONS FOR INSTALLING ELECTRICAL EQUIPMENT IN YOUR AREA Powering Up The Series 1 has a universal power supply that automatically adjusts to line voltages from 90 to 260 VAC After making electrical connections including grounding as described in Chapter 1 Installation of the Startup Guide press the power switch ...

Page 22: ...exposure is unavoidable either disconnect the oxygen cell from the Series 1 or equip the sample system with a valve that allows the cell to be switched to purge gas If the meter passes its self test the screen begins displaying measurements in a matrix format similar to the one shown in Figure 2 1 below Figure 2 1 Screen Matrix Format After Power Up HELP SELECT SETTINGS OPTIONS 1 Dew Point 2 Tempe...

Page 23: ...he user program SELECT lets you select the type of measurement and units to display SETTINGS enables you to set up recorders and alarms as well as enter probe data for basic operation To enter this menu you must enter the passcode See page 2 6 OPTIONS enables you to set up the screen to display measurements in text or graphical form It is also used to perform advanced functions You can use the scr...

Page 24: ...re the corresponding menu keys refer to Figure 2 3 Each menu key corresponds to each of the menu options in the user program Up to four new menu options appear each time you enter a menu The line above the menu options is the message line see Figure 2 3 below The message line displays the menu title the time and the date The message line also displays a list of selections depending on the menu opt...

Page 25: ...location for data entry Use the left arrow key as a backspace during number entry The response keys perform three functions Use either the YES or NO key to respond to questions Use the NO key to erase data Press YES to confirm a number entry or after erasing data to retrieve the previous number Entering the Passcode The Settings Menu is the only menu that requires a passcode The passcode is a four...

Page 26: ...r The NEXT menu option brings you to the following page and the PREV menu options brings you to the previous page Press DONE to exit and return to the current menu Figure 2 4 HELP Screen By pressing the ABOUT key users can access the history of their particular Series 1 its serial number PCI number software revision number and date of manufacture This information remains part of the HELP file and ...

Page 27: ...bration data for all necessary probes is properly entered in the System Calibration Menu as described on page 2 12 Verify that applicable high and low reference values for the Series 1 measurement circuitry are properly entered into the Reference Menu as described on page 2 21 Caution All high and low reference values are factory set and normally do not need adjustment However the factory may inst...

Page 28: ... to Entering Calibration Data on page 2 12 Verifying Probe Configuration Data IMPORTANT You should record this data on the Program Information List supplied in Appendix A of the Startup Guide or on a separate sheet of paper and keep it in a safe place 1 Enter the Probe Configuration Menu for a menu map refer to Figure 3 19 on page 3 32 A screen similar to Figure 2 5 below appears The Settings Menu...

Page 29: ...rent or voltage the Series 1 defaults to current Note If you are using an auxiliary input to measure pressure activate Au1 or Au2 depending on which input you are using in the pressure column for the desired channel then activate current or voltage in the Aux 1 or Aux 2 column In addition if you connect a pressure transducer to the PRESS OTHER terminal block you must activate the TF probe in the p...

Page 30: ...r a constant for moisture Kh temperature Kt or pressure Kp you must enter a constant value in the User Constant Menu See Entering Constants and User Functions on page 3 9 for instructions 4 Move the brackets to the correct probe type then press YES The selected probe type displays 5 Repeat Steps 1 through 4 for each channel 6 To exit press the DONE menu key until Main Menu appears on the message l...

Page 31: ...ble 2 3 below see Figure 3 19 on page 3 32 for a menu map The Settings Menu will prompt for a passcode see page 2 6 Table 2 2 Probe Types Measurement Mode Probe Type OXY Oxygen None Percent ppM Parts per Million ppB Parts per Billion HYG Hygrometry None MIS MIS Probe TF Three Function Probe Mxx M Series Probe Kh Constant Dew Point T Temperature None MIS MIS probe TF Three Function probe Mxx M Seri...

Page 32: ...m in a safe place If they are lost contact GE Panametrics for a duplicate If you alter any of the calibration data make sure you make the change on the Calibration Data Sheet Entering Moisture Probe Calibration Data You only need to enter calibration data for M and TF Series probes It is not necessary to enter calibration data for the Moisture Image Series Probe unless you send the probe back to G...

Page 33: ...YES 4 Enter the probe serial number from the Calibration Data Sheet and press YES The serial number is also scribed on the hex nut of the moisture probe The pointer automatically jumps to ND Press YES again 5 Enter the number of data points indicated on the Calibration Data Sheet typically 2 16 points for each probe and press YES The pointer automatically jumps to the first data point 6 Move the p...

Page 34: ...asure in ppm or percent Select the proper oxygen cell units in the Probe Configuration Menu The units you select will automatically appear in the second column Procedure for Entering Delta F Cell Calibration Data Referring to the Calibration Data Sheets enter calibration data separately for each probe on the designated channel as described below Figure 3 19 on page 3 32 for a menu map IMPORTANT St...

Page 35: ...urrent Multiplier at 1 00 unless you are using a background gas other than nitrogen See the section Background Gas Correction Factors for the Delta F Oxygen Cell in Chapter 2 of the Service Manual Repeat Steps 1 through 9 to enter oxygen calibration data for the remaining channels Figure 2 8 System Calibration Menu for Oxygen Cell To enter calibration data for other probes refer to Entering Moistu...

Page 36: ...MPORTANT Staple Calibration Data Sheets to the Program Information List Appendix A of the Startup Guide and store them in a safe place If they are lost contact GE Panametrics for a duplicate If you alter any of the calibration data make sure you make the change on the Calibration Data Sheet Note If you want to enter a constant value refer to Entering Constants and User Functions on page 3 9 It is ...

Page 37: ...from the Calibration Data Sheet and press YES 5 Move the pointer to the Zero mV mA line and press YES 6 Enter the voltage or current value and press YES 7 Move the pointer to the Zero PSIg line and press YES 8 Enter the PSIg value and press YES 9 Use Steps 5 through 8 to enter the span values Repeat steps 1 through 9 to enter pressure calibration data for the remaining channels To enter calibratio...

Page 38: ...be configuration see page 2 9 and a corresponding scale value Use the procedure below to enter data Figure 3 19 on page 3 32 for a menu map IMPORTANT Enter the auxiliary calibration data on the Program Information List Appendix A of the Startup Guide and store it in a safe place If it is lost contact GE Panametrics for a duplicate 1 At the System Calibration Menu see Table 2 3 on page 2 12 press t...

Page 39: ...ess YES to edit the Units label used when displaying the Auxiliary Input as described in Making Tags on page 2 38 7 Move the pointer to the first data point in the mA or VDC column and press YES The units for this column will depend on the units you selected in Activating and Changing Probes on page 2 9 8 Enter the voltage or current value and press YES 9 Move the pointer to the Units column and p...

Page 40: ...he reference data for that channel Table 2 4 shows the key sequences for entering the Reference Menu Figure 3 19 on page 3 32 for a menu map The Settings Menu will prompt for a passcode see page 2 6 Use the appropriate section that follows to verify or enter reference values for the probes Once you verify and or enter reference data for each input refer to Displaying Measurements on page 2 28 IMPO...

Page 41: ...Menu see Table 2 4 on page 2 21 press the PROBE menu key until the Moisture Reference Table appears see Figure 2 11 below Figure 2 11 Reference Menu for Moisture 2 Compare the data on the screen to the reference data printed on the label placed on the side or front of the unit or supplied with a replacement channel card If the data is incorrect use the procedure below to correct it If the data is ...

Page 42: ... steps 3 through 6 to enter the high and low reference values for the remaining channels Note You may also enter the Auto Cal Interval while in the Reference Menu Simply press the INTERVAL menu key and enter the desired interval See page 3 7 for more information To enter reference values for oxygen cell and pressure inputs refer to the following sections To exit press the DONE menu key until Main ...

Page 43: ...xygen Cell Reference Table appears see Figure 2 12 below Figure 2 12 Reference Menu for the Delta F Cell 2 Compare the data on the screen to the reference data printed on the label placed on the side or front of the unit If the data is incorrect use the following procedure to correct it If the data is correct do one of the following continue to enter reference values for moisture and pressure inpu...

Page 44: ...channel and press YES 7 Enter the zero value for the low reference for that channel and press YES 8 Move the pointer to the Span line for the low reference for the desired channel and press YES 9 Enter the span value for the low reference for that channel and press YES Repeat steps 3 through 9 to enter the high and low reference values for the remaining channels To enter reference values for moist...

Page 45: ...Figure 2 13 below Figure 2 13 Reference Menu for Pressure 2 Compare the data on the Series 1 screen to the reference data for the pressure input If the data is incorrect use the following procedure to correct it If the data is correct do one of the following continue to enter reference values for moisture and oxygen inputs as described in Entering Moisture Reference Values on page 2 22 and Enterin...

Page 46: ...lue for that channel and press YES 7 Repeat steps 3 through 6 to enter the high and low reference values for the remaining channels To enter reference values for moisture and oxygen inputs refer to Entering Moisture Reference Values on page 2 22 and Entering the Delta F Oxygen Cell Reference Values on page 2 25 To exit press the DONE menu key until Main Menu appears on the message line Then refer ...

Page 47: ...4 You can program each box to display any measurement for any channel Figure 2 14 Matrix Format Screen Each box in the matrix format is either assigned or unassigned An assigned box displays a channel number measurement mode units and a value An unassigned box displays no data See Figure 2 15 on page 2 29 for examples of each type of box HELP SELECT SETTINGS OPTIONS Main Menu 11 35 01 22 2000 pg 1...

Page 48: ... modes and units as they are displayed on the screen An example of a measurement is shown in Figure 2 16 on the next page When selecting a measurement mode please remember that some measurement modes require multiple inputs For example to display ppmv you need moisture and pressure inputs Table 2 5 below shows a list of measurement modes that require multiple inputs and what you need to program th...

Page 49: ...s a constant is being used in the measurement and an E indicates the meter is using Enhanced Response to determine the measurement An example of a measurement using a constant and Enhanced Response activated is shown in Figure 2 16 below The value is expressed in the units selected for a desired measurement mode Figure 2 16 Example of a Displayed Measurement 123 45 PPMw E K 5 H2O Value Indicators ...

Page 50: ...e H2O MH H DVM Digital Voltmeter Diagnostic Mode Moisture DVM VDC FH FH Diagnostic Mode MIS Probe FH Temperature T C Degrees Celsius default Temperature C T F Degrees Fahrenheit Temperature F T K Kelvin Temperature K T DVM Digital Voltmeter Diagnostic Mode Temp DVM VDC Pressure PSIg Pounds per Square Inch Gauge default Pressure PSIg Bars Bars Pressure Bars mbs Millibars Pressure mbs mm Hg Millimet...

Page 51: ...ually or automatically scroll through each page Switching from a Line Graph to the Matrix Format At the Main Menu press the keys as shown in Table 2 7 below Figure 3 18 on page 3 31 for a menu map The screen appears similar to Figure 2 17 below Press the DONE menu key until Main Menu appears on the message line This menu key only appears if you are displaying a line graph Figure 2 17 Matrix Screen...

Page 52: ...he PLACE menu key to move the larger pointer to the box you want to change 5 Move the brackets to the desired measurement mode and press the UNITS menu key A list of measurement units appears on the message line Note To clear unassign a box press NO The Series 1 labels the box X Unassigned see Figure 2 17 6 Move the brackets to the desired measurement units and press YES 7 Enter the desired channe...

Page 53: ...age number and press YES 3 At the Main Menu press the OPTIONS menu key 4 Press the SCREEN menu key 5 Press the CYCLE menu key 6 Enter an interval between 0 and 5 minutes and press YES Note Enter 0 00 to turn the automatic scroll off 7 Press the DONE menu key until Main Menu appears on the message line After you enter the interval the Series 1 begins scrolling through each page It will automaticall...

Page 54: ...ate section that follows to change the mode units or scale If you do not want to change any of the line graph settings press the DONE menu key until Main Menu appears on the message line This menu key will not appear if your screen is already set up for the graph format If you switched your screen to display a line graph earlier and are now returning to adjust the graph press the MATRIX menu key i...

Page 55: ...key 2 The Series 1 prompts for the minimum Y axis Enter the minimum value to be displayed and press YES 3 The meter then prompts for the maximum Y axis Enter the maximum value to be displayed and press YES 4 Press the X AXIS horizontal axis menu key 5 Enter the time in minutes and press YES The Series 1 accepts any value from 1 to 2160 minutes 36 hours 6 Press the DONE menu key until Main Menu app...

Page 56: ...menu key A list of units for the selected measurement mode appears on the message line 5 Select the desired units and press YES 6 Press the DONE menu key until Main Menu appears on the message line Note You can rescale graph data at any time without losing data by entering new X axis and Y axis values Tagging the Inputs The Series 1 enables users to assign tags to customize the display of the inpu...

Page 57: ...Edit Box and four menu options Figure 2 19 Label Entry Menu 3 A flashing select pointer appears in the upper left corner of the character set Use the NEXT ROW and PREV ROW menu keys to move the pointer to the desired row 4 Then use the arrow keys to move the pointer to the desired character and press YES The selected character will appear in the Edit Box The upper left corner of the character set ...

Page 58: ...bel press the DONE menu key The Probe Configuration Menu screen appears again this time with a tag symbol appearing beside the tagged probe When you return to the Main Menu the matrix will display the tag name in the assigned channel Note Tags do not appear for alarm or recorder outputs ...

Page 59: ...small changes To exit press the DONE menu key until Main Menu appears on the message line Figure 2 20 Contrast Adjust Menu Note The screen in Figure 2 20 above will appear in either the matrix format or a line graph depending on the screen setup Table 2 10 Moving from Contrast Adjust to Main Menu Press the following menu keys To enter the OPTIONS Option Menu SCREEN Screen Menu CONTRAST Contrast Ad...

Page 60: ...igits for the month day year hours 0 23 minutes and seconds If you do not want to change a part of the time or date press the pointer to skip over it until you complete entering the year 3 To exit press the DONE menu key until Main Menu appears on the message line Figure 2 21 Setting the Clock Calendar Table 2 11 Moving from System Service to Main Menu Press the following menu keys To enter the SE...

Page 61: ......

Page 62: ...Your passcode is 2719 Please remove this page and put it in a safe place for future reference ...

Page 63: ...Chapter 3 ...

Page 64: ...etting Up the Alarms 3 4 Using Automatic Calibration 3 7 Entering Constants and User Functions 3 9 Using Computer Enhanced Response 3 15 Setting Up the Data Logger 3 17 Setting Up the RS232 Communications Port 3 22 Setting Error Processing 3 28 Loading New Software 3 30 ...

Page 65: ...o select the measurement mode units and range for each recorder press the menu keys as shown in Table 3 1 below The screen appears similar to Figure 3 1 on page 3 2 Figure 3 19 on page 3 32 for a menu map IMPORTANT Switch Blocks must be set to the appropriate positions for output ranges See Connecting the Recorder Outputs in Chapter 1 of the Service Manual To test or trim recorders refer to Testin...

Page 66: ...screen The screen will only display installed channels 2 Press the SET menu key A box highlights Recorder A and the menu keys change to RANGE MODE and A B 3 Press the A B menu key to move the box to the desired recorder 4 Press the MODE menu key A list of available measurement modes appears on the message line Refer to Table 2 6 on page 2 31 for a list of measurement modes 5 Move the brackets to t...

Page 67: ... brackets to the desired output range and press YES 4 With the arrow key move the pointer to the Zero line for the desired recorder Press YES to erase the present value 5 Enter a new value and press YES 6 Move the pointer to Span for the desired recorder and press YES to erase the current value 7 Enter a new value and press YES 8 Repeat steps 1 through 6 to set up the other recorder Repeat all of ...

Page 68: ...ng and Maintenance of the Service Manual The Settings Menu will prompt for a passcode see page 2 6 Figure 3 2 Alarm Settings Menu Note The first time you enter this menu all alarm relays are OFF When an alarm value is not entered the alarm relay is OFF Table 3 2 Moving to Alarm Settings Menu Press the Main Menu key To enter the SETTINGS Settings Menu OUTPUTS Output Settings Menu ALARMS Alarm Setti...

Page 69: ...e the brackets to the desired units and press YES 6 Press the TRIP menu key The pointer will move to the Trip line Use the pointer to select Above or Below and thus determine whether the alarm is a High or Low alarm Then press YES and then the DONE key Entering the Setpoint and Deadband Values for Alarms Figure 3 19 on page 3 32 for a menu map IMPORTANT Select the measurement mode and units before...

Page 70: ...s outside the deadband range On a High alarm the alarm will trip if the measurement is greater than or equal to the setpoint It will not RESET until the measurement is less than the setpoint minus the deadband However on a Low alarm the alarm will trip if the measurement is less than or equal to the setpoint It will not RESET until the measurement is greater than the setpoint plus the deadband 5 M...

Page 71: ...nditions Note If you are using only Moisture Image Series Probes to measure moisture you do not have to Auto Cal To set the Auto Cal interval press the keys shown in Table 3 3 The screen appears similar to Figure 3 3 below Refer to Figure 3 19 on page 3 32 for a menu map The Settings Menu will prompt for a passcode see page 2 6 Figure 3 3 Setting the Auto Cal Interval Table 3 3 Entering the Refere...

Page 72: ...nu appears on the message line The next time Auto Cal occurs will depend on the length of the time interval that was set See the example below EXAMPLE The Series 1 establishes a fixed schedule beginning at midnight using the interval specified to determine the times of subsequent Auto Cals For example if you enter a 1 hr 30 min time interval Auto Cal will occur 16 times per day 1 day 1440 minutes ...

Page 73: ...er of the box in matrix format for channels that are using a user constant value A symbol does not appear in the graph format The other type of constant is a saturation constant The Series 1 requires a saturation constant in order to calculate ppmw in non aqueous liquids If you do not know the saturation constant of the liquid contact GE Panametrics To extend programming capability user functions ...

Page 74: ...e Program Information List supplied in Appendix A of the Startup Guide or on a separate sheet of paper and keep it in a safe place 1 After entering the User Constant Menu refer to Table 3 4 on page 3 9 press the PAGE menu key until the User Constant Table appears as shown in Figure 3 4 below 2 Move the pointer to the desired channel and constant to enter and press YES to erase the present value 3 ...

Page 75: ... in a new entry and press YES to confirm the entry 3 Move the pointer to the Decimal Places line Press YES to erase any current number and enter the number of decimal places to the right of the decimal point that you wish to appear Press YES to confirm the entry 4 Move the pointer to the Units Symbol line Press YES to remove any entry type in a new entry and press YES to confirm the entry 5 Move t...

Page 76: ...list of parameters will appear Scroll through the list until you reach the desired parameter and press YES 5 A list of units will then appear Scroll through the list with the arrow keys until you reach the desired units and press YES Note If you wish to link a parameter from a particular channel add Ch after you select the parameter For example DP C 2 is the dewpoint in C on channel 2 6 Press the ...

Page 77: ...ss YES Enter the number of data points desired and press YES to confirm the entry 2 The pointer moves to the Title entry Press YES and the Label Entry Menu shown in Chapter 2 appears Note The title is for information purposes and does not appear anywhere else 3 A flashing select pointer appears in the upper left corner of the character set Use the NEXT ROW and PREV ROW keys to move the pointer to ...

Page 78: ... 4 on page 3 9 press the PAGE menu key until a screen similar to Figure 3 8 appears Figure 3 8 Cs Constant Menu 2 Use the CHANNEL menu key to cycle to the desired channel 3 Use the arrow keys to go to the first data point for temperature and press YES 4 Enter the temperature for the first data point and press YES Then press the pointer to move to Cs 5 Press YES to erase the current value 6 Enter t...

Page 79: ... a reasonably constant final dew point and flow rate have a minimum flow rate of one standard cubic foot per hour SCFH operate at atmospheric pressure 0 psig have an actual moisture content at or above a dew frost point of MIS Probe 85 C M Series or TF Series Probe 110 C To activate and use enhanced response see Figure 3 18 on page 3 31 for a menu map and complete the following steps 1 Allow the s...

Page 80: ... to dry gas IMPORTANT You must enable enhanced response before you expose the sensor to the dry gas The meter must see all of the dry down in order to make accurate calculations When activating enhanced response for a channel a reverse video E symbol appears in the top right hand corner of the corresponding box when displaying data in the matrix format When the meter determines the final value the...

Page 81: ...s and minutes the log will run before the programmed End Time see Figure 3 10 The meter assigns logged data a record and a header The record consists of the date time and corresponding logged values The header consists of the selected channels the selected measurement modes and the time the log started It is used to distinguish one log from another The Series 1 can store logs in six slots Thus if ...

Page 82: ...ous pages of logged data To view other records continue to press the NEXT menu key until end appears at the end of the record If the screen has more than three columns use the pointer to scroll across the columns To exit press the DONE menu key until Main Menu appears on the message line Removing Logged Data To remove a log select it with the arrow keys on the Data Logger Menu Press the FILE menu ...

Page 83: ... Label Entry Menu shown in Figure 3 12 below Figure 3 12 Label Entry Menu 2 Specify the log name Use the NEXT ROW and PREV ROW menu keys to scroll to the desired line and then use the arrow keys to reach the desired letters numbers or symbols When you have completed the label press the DONE menu key The Set Data Logger Menu now appears as shown in Figure 3 13 below Figure 3 13 Set Data Logger Menu...

Page 84: ...second HH MM SS format for example 15 33 00 Press YES when you have completed entering the data for each time 6 The pointer then moves to the Interval row Specify the log interval in minutes and seconds The interval can be as short as 5 sec Skip steps 7 through 9 and proceed to Selecting the Channel Measurement Mode and Units on the next page Entering Settings for a Cyclic Log The screen now appea...

Page 85: ...interval times the number of records Selecting the Channel Measurement Mode and Units 1 The entry for each log channel and mode appears as ___ ____ if blank Use the arrow keys to move to the entry to change or enter data Press NO to erase a previous entry 2 Press numeric keys 1 to 6 to select the desired channel 3 Press the MODE menu key A list of measurement modes appears on the message line see ...

Page 86: ...age 3 32 for a menu map A screen appears similar to Figure 3 15 below The Settings Menu will prompt for a passcode see page 2 6 Figure 3 15 Port Settings Menu Use the steps that follow to select the function baud rate report format and interval for the serial port Other serial port settings are set and cannot be changed The fixed settings are as follows 8 bit word length no parity 1 stop bit flow ...

Page 87: ... data Note If you are using GE Panametrics PanaLink software set the baud rate to 19 200 4 The next step will depend upon the function selected in Step 1 a If you selected PanaLink as the function press the NODE ID menu key Unless you are instructed to change the ID number by GE Panametrics please leave the number at the default ID of 16 Press YES to confirm the entry b If you selected a printer a...

Page 88: ... YY format The Series 1 uses a zero as a column place holder so the date will always be eight characters long The date is always followed by one ASCII space Time The time of the report is shown in HH MM SS format Time is in 24 hour format The Series 1 uses a zero as a column place holder so the time will always be eight characters long The time is always followed by one ASCII space Channel The cha...

Page 89: ...ary with the magnitude and type of data see Table 3 9 on page 3 26 for number formats for each mode units Signed values will contain a leading plus or minus sign Unsigned values percent oxygen for example have no sign The data is always followed by two ASCII spaces CR LF Every record is terminated with a carriage return line feed pair ASCII 13 and 10 decimal 0D and 0A hexadecimal ...

Page 90: ...llion of water by weight MCF IG X XXXX Pounds of water per million standard cubic feet of an ideal gas MCF NG X XXXX Pounds of water per million standard cubic feet of natural gas ppMv NG XXX XXX Parts per million of water by volume in natural gas MH X XXXX GE Panametrics moisture calibration unit for M and TF Series probes FH X XXXX GE Panametrics moisture calibration unit for Moisture Image Seri...

Page 91: ...re in kiloPascals absolute P mV XXX XX Pressure transducer output in millivolts FP XXXX XX GE Panametrics pressure calibration unit in millivolts P DVM X XXXX Pressure circuit voltage in volts DC Auxiliary Inputs Aux1 V X XXXX Auxiliary input 1 in volts Aux1 I XX XXX Auxiliary input 1 in milliamperes Aux1 F XX XXXX Auxiliary input 1 in user defined units Aux2 V X XXXX Auxiliary input 2 in volts Au...

Page 92: ...Ignore this forces the recorder analog outputs to 0 mA or 0V Procedure for Setting Error Processing The Series 1 user program allows you to turn Error Messaging either ON or OFF In addition if Error Messaging is ON you can program separate Over Range and Under Range responses for all alarm relays if installed and recorder analog outputs The error responses are independently programmable for each c...

Page 93: ...urs Recorders Analog Outputs The available actions are Ignore High Low Hold last value and Value specified by user For the Ignore option both the A and B analog output signals are forced to 0 mA or 0 V when a range error occurs 5 Repeat Steps 2 4 for both ranges and for each output on the desired channel s To exit Error Processing press the DONE menu key until Main Menu appears on the message line...

Page 94: ...ations Port in Chapter 3 of this manual and Connecting a PC or Printer in Chapter 1 of the Service Manual 4 Start the communications program on the PC and select the COM port with the connection to the MIS 1 5 Set the following information Baud Rate 19200 Data Bits 8 Parity none Stop Bits 1 Flow Control none 6 Turn on the power to the MIS 1 7 Press and hold the 0 key Note The display will indicate...

Page 95: ...u See Figure 3 18 SCREEN USER LOGGER DONE If you are displaying data in Matrix Format or Line Graph GRAPHS CYCLE CONTRAST DONE Matrix Screen Menu MATRIX SCALE CONTRAST DONE Graph Screen Menu HELP DECR INCR DONE Contrast Adjust Menu HELP DONE HELP X AXIS Y AXIS DONE Graph Scale Menu HELP DECR INCR DONE Contrast Adjust Menu HELP STORAGE FILE DONE Data Logger Menu PAGE HELP TABLE DONE User Constant M...

Page 96: ...Menu SET BAUD FUNCTION INTERVAL DONE Port Settings Menu RANGE MODE DONE Set Recorder SET TEST CHANNEL DONE Recorder Settings Menu TRIM DONE Recorder Test Menu HELP HELP Help System SELECT Display Select Menu See Figure 3 17 See Figure 3 17 DONE SERVICE HELP REFS CLOCK DONE System Service Menu HELP INTERVAL PROBE DONE Reference Menu OPTIONS Option Menu See Figure 3 17 HELP TAG DONE Probe Configurat...

Page 97: ...Chapter 4 ...

Page 98: ...n 4 1 Changing Measurement Modes and Units 4 1 Changing a Matrix Display to a Graph 4 6 Changing Alarm Settings 4 7 Changing Recorder Settings 4 9 Programming User Functions 4 12 Entering User Tables 4 14 Entering Saturation Constants 4 15 ...

Page 99: ...of PanaView s capabilities see the PanaView User s Manual IMPORTANT Before attempting to display or change programming via the PC be sure to install PanaView as described in Chapter 1 of the PanaView User s Manual Changing Measurement Modes and Units PanaView enables you to actually change the measurements on the matrix display For example through the PC you can switch a window on page 2 from its ...

Page 100: ... shown in Figure 4 1 on page 4 3 Each individual blue window shows the mode and unit measured in the corresponding window on the Series 1 page with the assigned channel number in the upper left corner A line below indicates which page the meter is currently displaying 5 To scroll to a particular page click on the left or right pointer buttons at the bottom of the window 6 If you want to set a part...

Page 101: ...Programming with PanaView 4 3 April 2004 Open the Meter Display Window cont Figure 4 1 The Meter Display Window ...

Page 102: ...he unit and the Series 1 reassigns the selected box on the Series 1 and in the Meter Display window c To unassign a mode and unit without assigning a different sensor click on the desired window and then on Unassign Figure 4 2 on page 4 5 shows a typical change on the PC screen from O2 ppm shown in Figure 4 1 on page 4 3 to PSIg Note For an explanation of unit abbreviations see Table 2 6 on page 2...

Page 103: ...Programming with PanaView 4 5 April 2004 Change Meter Programming cont Figure 4 2 Matrix Display After Programming ...

Page 104: ...nd units 3 Double click on the desired unit and enter the parameters for the graph as shown in Figure 4 3 below 4 Click on Apply to change the graph and on OK to confirm the change and close the window 5 Click on the Set Active Display button to prompt the Series 1 screen to display the graph To toggle between the matrix display and the graph click on the Matrix Display or Graph Display buttons Th...

Page 105: ...1 From the New Meter Browser option pull down the desired meter and click on the desired channel 2 From the channel menu click on the Alarms option and then on Alarm A or Alarm B The Alarm window appears as shown in Figure 4 4 below 3 To set the alarm sensor and unit a Pull down the Sensor list box in the lower part of the window b Then click on the desired sensor type c A list of units appears Cl...

Page 106: ...arm will react when it encounters an error a Pull down the High list box under the On Error line b Then click on the desired option from the listing Ignore error Trip on error Reset on error or Hold on last error c Repeat the procedure for the Low list box 7 To test the alarm click on the Test button Click on the Trip button to trip the alarm and on the Reset button to reset it 8 Click on Apply to...

Page 107: ...ppears as shown in Figure 4 5 below 3 To set the recorder type pull down the Type list box and click on one of the three choices 0 to 20 mA 4 to 20 mA and 0 to 2 V 4 To set the recorder unit pull down the Sensor Class list box and click on the desired sensor Then pull down the Unit list box and select from the available units 5 To establish the input settings enter the desired values in the Zero a...

Page 108: ...own in Figure 4 6 below At the top slide the pointer to the desired percentage of the full scale recorder range and click on OK 3 To trim the recorder outputs scroll the arrows to the right of the Zero and Span windows up or down until you reach the desired values and click on OK Note For a complete discussion of testing and trimming recorder outputs refer to the sections Testing the Recorder Outp...

Page 109: ... 3 To program the meter s response to errors pull down the On Low Error list box and click on one of five options Ignore error Go to high limit Go to low limit Hold last value Go to error value 4 Repeat this selection with the On High Error list box 5 Click on Apply to apply the changes immediately and on OK to confirm the changes and close the window To return the window to its previous settings ...

Page 110: ...C To program a user function 1 From the New Meter Browser option click on the desired communications port meter and channel 2 Click on User Functions and then double click on the desired function 1 to 4 The right side of the screen appears similar to Figure 4 8 below 3 Enter three parameters a function label of no more than 16 characters a function symbol no more than 4 characters and the number o...

Page 111: ...ly or to erase incorrect values or operations 6 When you have completed entering the function click on Apply to apply the function immediately and on OK to confirm the function and close the window To return the window to its previous settings click on the Cancel button Note It may be advisable to use the Apply button as a means of checking the function s validity If the function is improperly typ...

Page 112: ...able 1 From the New Meter Browser option click on the desired communications port and meter 2 Click on User Tables and then double click on the desired table The right side of the screen appears similar to Figure 4 9 Figure 4 9 User Table Window 3 Enter a name up to 13 characters for the table in the Table Name text box and enter up to 16 X and Y values in the appropriate columns The table now app...

Page 113: ...1 below Figure 4 11 Constant Saturation Window 4 Enter up to six data points in the text boxes to establish a curve of saturation constants versus temperature 5 The window also allows you to specify whether you wish Enhanced Response for the channel Enhanced Response uses a dynamic moisture calibration technique to extrapolate the moisture level to the end point when making measurements in abrupt ...

Page 114: ...es Probe 1 8 Calendar Setting 3 15 Calibration Data 2 13 2 17 Oxygen Cell 2 15 System Calibration Menu 2 12 Calibration Data Sheets 2 12 Changing Matrix to Graph Display in PanaView 4 6 Changing Recorder Settings in PanaView 4 9 Channel Line Graph Selecting 2 37 Matrix Format Selecting 2 33 Clock Setting 3 15 Communications Port 3 22 Default Settings 3 22 Interval Selection 3 23 Setting Up 3 22 Co...

Page 115: ...e in PanaView 4 15 Enhanced Response Menu 3 15 Entering User Defined Tables 3 13 Erasing Assigned Boxes 2 33 Error Handling in PanaView 4 11 Error Processing 3 28 F Functions 2 6 Functions Screen and Key 2 5 Functions User 3 11 G Graph Screen Menu 2 35 H Header Definition 3 17 Help Getting On Line 2 7 I Indicator Definition 2 29 Inputs Labeling 2 12 Inputs Tagging 2 37 K Key Functions 2 5 2 6 Arro...

Page 116: ... Enhanced Response Menu 3 15 Graph Screen Menu 2 35 Help Menu 2 7 Main Menu Descriptions 2 4 Port Settings Menu 3 22 Probe Configuration Menu 2 9 Recorder Settings Menu 3 1 Reference Menu 2 21 System Calibration Menu 2 12 System Service Menu 3 15 User Constant Menu 3 9 User Constant Menu Enhanced Resp 3 15 Message ID 3 24 Mode Unit Sample Output 3 24 Moisture Image Series Probe 1 6 Description of ...

Page 117: ...Cables 1 8 Calibration Data Entering 2 17 Reference Values Entering 2 26 Probe Configuration Menu 2 9 Probes Activating Changing in User Program 2 9 Calibration Data Sheets 2 12 Calibration Data Entering 2 12 Moisture Probes 1 4 Oxygen Cells 1 4 Pressure Transducers 1 4 Pressure Transmitters 1 4 Programming User Program 1 9 Programming Error Handling in PanaView 4 11 Programming User Functions in ...

Page 118: ... Menu 2 9 Recorders Settings Menu 3 1 Reference Menu 2 21 System Calibration Menu 2 12 System Service Menu 3 15 Software Loading 3 30 Specifications Cables 1 8 Specifying User Functions 3 11 System Service Menu 3 15 T Tables Entering in PanaView 4 14 Tables User Defined 3 13 Tagging Inputs 2 37 Testing Recorder Outputs in PanaView 4 10 TF Series Probe Cables 1 8 Description of 1 4 1 5 Top Drain Ce...

Page 119: ...n accordance with all appropriate international national and local standard codes and practices and site regulations for flameproof apparatus and in accordance with the instructions contained in the manual Access to the circuitry must not be made during operation Only trained competent personnel may install operate and maintain the equipment The product has been designed so that the protection aff...

Page 120: ...td EECS Buxton SK17 9JN UK EN 61326 1998 Class A Annex A Continuous Unmonitored Operation EN 61010 1 1993 A2 1995 Overvoltage Category II Pollution Degree 2 following the provisions of the 89 336 EEC EMC Directive the 73 23 EEC Low Voltage Directive and the 94 9 EC ATEX Directive The units listed above and any sensors and ancillary sample handling systems supplied with them do not bear CE marking ...

Page 121: ...ass A Annex A Continuous Unmonitored Operation EN 61010 1 1993 A2 1995 Overvoltage Category II Pollution Degree 2 suivant les régles de la Directive de Compatibilité Electromagnétique 89 336 EEC de la Directive Basse Tension 73 23 EEC et d ATEX 94 9 EC Les matériels listés ci dessus ainsi que les capteurs et les systèmes d échantillonnages pouvant être livrés avec ne portent pas le marquage CE de ...

Page 122: ...Buxton SK17 9JN UK EN 61326 1998 Class A Annex A Continuous Unmonitored Operation EN 61010 1 1993 A2 1995 Overvoltage Category II Pollution Degree 2 gemäß den Europäischen Richtlinien Niederspannungsrichtlinie Nr 73 23 EG EMV Richtlinie Nr 89 336 EG und ATEX Richtlinie Nr 94 9 EG Die oben aufgeführten Geräte und zugehörige mitgelieferte Sensoren und Handhabungssysteme tragen keine CE Kennzeichnung...

Page 123: ... 90 E mail panametrics panametrics fr Germany Mess und Pruftechnik Robert Bosch Straße 20a 65719 Hofheim Germany Telephone 49 6122 8090 Fax 49 6122 8147 E mail panametrics t online de Italy Via Feltre 19 A 20132 Milano Italy Telephone 02 2642131 Fax 02 26414454 E mail info panametrics it Japan 2F Sumitomo Bldg 5 41 10 Koishikawa Bunkyo Ku Tokyo 112 0002 Japan Telephone 81 03 5802 8701 Fax 81 03 58...

Page 124: ...Telephone 781 899 2719 Toll free 800 833 9438 Fax 781 894 8582 E Mail panametrics ps ge com Web www gepower com panametrics Ireland GE Panametrics Shannon Industrial Estate Shannon County Clare Ireland Telephone 353 61 470200 Fax 353 61 471359 E Mail info panametrics ie ...

Page 125: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Reviews: