6 Alarms
167
ALARMS TEST
Alarms tests check the following alarms:
Tube Disconnect
Patient Disconnect
Sensor Disconnect (Open Loop)
AC Disconnect
Low O2 Pressure
Low Minute Volume
High Pressure
High O2
Low O2
Apnea
High PEEP
Auto Start
WARNING
Do not perform these tests while the patient is connected to the
machine.
To perform an alarm test:
1.
Connect the ventilator to AC power and oxygen supply according to the
iVent
TM
201 specification (See page 32).
2.
Power up the ventilator. It automatically runs a self-test.
3.
Select the default set for patient weight 70kg. Set FiO2 to 40% (if default set is
configured to a different value).
4.
Run O.V.T. (See page 60).
5.
Attach a 2L test lung with Rp20 resistance to the patient circuit through an
HME filter and start ventilation.
6.
Tube Disconnect
alarm test
. Wait for at least 30 seconds after the
ventilation starts, and disconnect the patient circuit from the ventilator
outlet port. Verify that the
Tube Disconnect
alarm is activated during the
first inhale after disconnection. Reconnect the patient circuit to the
ventilator and verify that the
Tube Disconnect
alarm disappears
automatically.
7.
Patient Disconnect alarm test
. Disconnect the test lung from the patient
circuit. Verify that the Patient Disconnect alarm is activated. In case a
remote alarm cable is connected, verify that alarm at remote station is also
activated.
Reconnect the test lung to the patient circuit, and verify that the
Patient
Disconnect
alarm disappears automatically.
Summary of Contents for iVent 201 Series
Page 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Page 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Page 4: ......
Page 14: ......
Page 42: ......
Page 68: ...56 Notes...
Page 70: ......
Page 98: ...86 Notes...
Page 100: ......
Page 140: ...128 Notes...
Page 142: ......
Page 152: ......
Page 154: ......
Page 184: ......
Page 201: ...7 Care Maintenance and Tests 189 Notes...
Page 202: ......
Page 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Page 204: ......
Page 210: ...198 Notes...
Page 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Page 212: ......
Page 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Page 216: ......
Page 219: ...Appendix C Operating Theory 207...
Page 220: ......
Page 221: ...Appendix C Operating Theory 209 Notes...
Page 222: ......
Page 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Page 224: ......
Page 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Page 234: ......
Page 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Page 236: ......
Page 243: ...Appendix E Ventilation Modes 231 Notes...
Page 244: ......
Page 246: ......
Page 252: ...240 Notes...
Page 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Page 254: ......
Page 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Page 278: ......
Page 280: ......
Page 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Page 282: ......
Page 286: ...274...
Page 289: ...Index 277...