Appendix G: Accessories
247
Instructions for use
1.
Connect the two channel flow sensor tubes to their respective
connectors next to the Breathing Gas Outlet. Flow Sensor tubes are
equipped with luer connectors - left female and right male (Figure 185
(B)) for differentiation.
2.
Connect the Exhalation Control Tube with luer connector to its respective
connector on the right side of the front panel (C).
3.
Perform the standard Operational Verification (O.V.T) test procedure, as
described on page 60, prior to use on a patient.
4.
Connect the Flow Transducer to the patient airway when ready.
NOTE
It is recommended that an HME filter be used when connecting to the
patient
Instructions for use with the humidifier
NOTE
For proper use and application of the Humidifier, refer to the
manufacturers operator instructions.
the P/N of the circuit with the humidifier configuration is M1161731.
1.
Connect the one-way valve to the ventilator outlet (Figure 178 (A))
2.
Connect 12-in. length of 22-mm tubing to one-way valve.
3.
Connect opposite end of tubing to humidifier inlet, and connect 22-mm
tubing (with the installed watertrap) between humidifier outlet and Y-
connector (Figure 178 (F)).
4.
If using Patient circuit with heated wire: Connect full-length tubing with
heated wire to Y-connector (Figure 178 (F)). Note: The P/N of the circuit
with the heated wire configuration is M1161732
5.
Connect the Exhalation valve (Figure 178 (D)) to the Y-connector (F) with
22-mm tubing (with the installed watertrap). Verify that the arrow on the
exhalation valve cup is directed to opposite from the patient.
6.
Connect the two channel flow sensor tubing (Figure 178 (B)) to their
respective connectors next to the Breathing Outlet. Flow Sensor tubing
are equipped with luer connectors - left female and right male for
differentiation.
7.
Connect the Exhalation Control Tube (Figure 178 (C)) with luer connector
to its respective connector on the right side of the front panel, marked
with blue dot.
8.
Perform the standard Operational Verification (O.V.T) test procedure, as
described on page 60, prior to use on a patient.
9.
Connect the Flow Transducer to the patient airway when ready.
NOTE
Always position the Water trap so that it sits below the patient circuit and
Summary of Contents for iVent 201 Series
Page 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Page 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Page 4: ......
Page 14: ......
Page 42: ......
Page 68: ...56 Notes...
Page 70: ......
Page 98: ...86 Notes...
Page 100: ......
Page 140: ...128 Notes...
Page 142: ......
Page 152: ......
Page 154: ......
Page 184: ......
Page 201: ...7 Care Maintenance and Tests 189 Notes...
Page 202: ......
Page 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Page 204: ......
Page 210: ...198 Notes...
Page 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Page 212: ......
Page 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Page 216: ......
Page 219: ...Appendix C Operating Theory 207...
Page 220: ......
Page 221: ...Appendix C Operating Theory 209 Notes...
Page 222: ......
Page 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Page 224: ......
Page 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Page 234: ......
Page 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Page 236: ......
Page 243: ...Appendix E Ventilation Modes 231 Notes...
Page 244: ......
Page 246: ......
Page 252: ...240 Notes...
Page 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Page 254: ......
Page 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Page 278: ......
Page 280: ......
Page 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Page 282: ......
Page 286: ...274...
Page 289: ...Index 277...