6 Alarms
163
Warning
Set Count or
Time of
Condition
Mode of Failure Priority Probable Cause Criterion for
Clearing Alarm
Corrective Action
Low Pressure
3 breath delay
(4 with setting
change)
Airway pressure
during inspiration
is lower than the
Low Pressure
alarm threshold
Mid
Clinical condition
Incorrect alarm
setting
Patient
disconnect or
high leak
Pressure is
higher than
alarm threshold
for 2 breaths
Subject to clinical
judgment, correct
setting as
necessary
Verify patient
circuit
Low SpO2
2 breaths
delay
SpO2 level is
below the set
lower SpO2 limit
High
The patient
clinical condition,
FiO
2
level set too
low, or SpO2
alarm setting too
high
SpO2 level is
higher than the
alarm limit from
1 breath.
Assessing the
patient clinical
condition,
increasing the
FiO
2
concentration, or
decreasing the
SpO2 alarm
setting.
Auto Start
Patient is
connected but
the unit is in
Standby mode
High
Mid
User enters
Standby mode
unintentionally
User connects
the ventilator to
the patient but
fails to press
Start
N/A Clear
the
alarm
and set the
required
ventilation
parameters
Service Notice Technical
irregularity
(needs Technical
Support Services)
High Power
surge
Multiple sensor
disconnection
Calibration
required
Ventilator
Malfunction
N/A
Turn the system
OFF and ON
again If the alarm
is not cleared,
remove the unit
for inspection and
servicing by
qualified
technical
personnel
Warm start-up
service notice
Software
detected
technical
irregularity
High Software
exception
N/A 1.
Single event –
no action
required –
resumed
ventilation with
previous
settings in up
to 30 seconds.
2.
Multiple
repeated event
– remove the
unit for
inspection and
servicing by
Summary of Contents for iVent 201 Series
Page 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Page 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Page 4: ......
Page 14: ......
Page 42: ......
Page 68: ...56 Notes...
Page 70: ......
Page 98: ...86 Notes...
Page 100: ......
Page 140: ...128 Notes...
Page 142: ......
Page 152: ......
Page 154: ......
Page 184: ......
Page 201: ...7 Care Maintenance and Tests 189 Notes...
Page 202: ......
Page 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Page 204: ......
Page 210: ...198 Notes...
Page 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Page 212: ......
Page 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Page 216: ......
Page 219: ...Appendix C Operating Theory 207...
Page 220: ......
Page 221: ...Appendix C Operating Theory 209 Notes...
Page 222: ......
Page 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Page 224: ......
Page 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Page 234: ......
Page 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Page 236: ......
Page 243: ...Appendix E Ventilation Modes 231 Notes...
Page 244: ......
Page 246: ......
Page 252: ...240 Notes...
Page 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Page 254: ......
Page 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Page 278: ......
Page 280: ......
Page 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Page 282: ......
Page 286: ...274...
Page 289: ...Index 277...