162
Warning
Set Count or
Time of
Condition
Mode of Failure Priority Probable Cause Criterion for
Clearing Alarm
Corrective Action
Patient Circuit
Failed
10 breath
delay
Insufficient
resistance from
one way valve
High
One way valve
has been
removed from
patient circuit or
has deteriorated
(ripped or torn or
missing
membrane)
Substitute one
way valve with
new one
Replace patient
circuit
(disposable) or
replace and/or
service one way
valve (re-useable)
Low Heart
Rate
2 breaths
delay
The heart rate is
below the rate
limit
High Clinical
Condition Heart rate
exceeds the
lower alarm
limit for 1
breath
Check the patient
condition.
Low O
2
15 breath
delay
Oxygen
concentration is
below alarm
setting
High (LT) Alarm set too
high
No or low oxygen
flow to system
Calibration
required
High variability in
patient breathing
F
i
O
2
is higher
than alarm
setting for 2
breaths
Check oxygen
supply
Verify delivered
O
2
concentration
with an external
analyzer
Refer unit to
qualified service
technician for
calibration
Decrease alarm
settings
Change from
Adaptive Flow™
to fixed manual
flow
Low O
2
Pressure*
10 seconds
after the
detection of O
2
inlet pressure
below 10 PSI
External oxygen
supply pressure
is low or
disconnected
High (LT) Oxygen supply is
disconnected
Low oxygen
pressure
Nebulizer is
activated without
Oxygen supply.
N/A
Adequate O
2
pressure is
detected
Reconnect or
replace oxygen
supply source
Ensure proper
regulation of
external oxygen
pressure
* While “Low O2 Pressure” alarm condition is active the FiO2 field on the Main screen flashes red
and instead of the set FiO2, the measured FiO2 value is displayed.
Summary of Contents for iVent 201 Series
Page 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Page 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Page 4: ......
Page 14: ......
Page 42: ......
Page 68: ...56 Notes...
Page 70: ......
Page 98: ...86 Notes...
Page 100: ......
Page 140: ...128 Notes...
Page 142: ......
Page 152: ......
Page 154: ......
Page 184: ......
Page 201: ...7 Care Maintenance and Tests 189 Notes...
Page 202: ......
Page 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Page 204: ......
Page 210: ...198 Notes...
Page 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Page 212: ......
Page 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Page 216: ......
Page 219: ...Appendix C Operating Theory 207...
Page 220: ......
Page 221: ...Appendix C Operating Theory 209 Notes...
Page 222: ......
Page 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Page 224: ......
Page 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Page 234: ......
Page 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Page 236: ......
Page 243: ...Appendix E Ventilation Modes 231 Notes...
Page 244: ......
Page 246: ......
Page 252: ...240 Notes...
Page 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Page 254: ......
Page 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Page 278: ......
Page 280: ......
Page 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Page 282: ......
Page 286: ...274...
Page 289: ...Index 277...