ADAPTIVE BI-LEVEL ALARM SETTINGS
The Default Alarm Settings for Adaptive Bi-Level are provided below (Table 7):
Table 7: Adaptive Bi-Level Default Alarm Settings
Rate
Default for High Rate is Off
Default for Low Rate is 6
Minute Volume
Default for Low Minute Volume set to 0
Default for High Minute Volume is based on weight
Pressure
Default for High Pressure is 20
Default for Low Pressure is 5
Apnea Time
Default for Apnea Time is 20 seconds
FiO
2
Default for High FiO
2
is +20% from set value
Default for Low FiO
2
is 21% or –10% from set value
Leak
Default for Leak is 100%
NOTE
Auto Settings are disabled in Adaptive Bi-Level
OTHER ALARM OPTIONS
In addition to the Alarm Settings the operator can disable the Inverse I:E Ratio
alarm and the patient circuit disconnect alarm.
Inverse I:E Ratio (ON – OFF option, default set to OFF)
Tidal Volume not Delivered (field deactivated)
Patient Circuit Disconnect (ON or OFF – default set to ON)
ADJUSTING ADAPTIVE BI-LEVEL PARAMETERS
Use the
Mode
screen to set parameters for Adaptive Bi-Level ventilation (Figure
127)
Summary of Contents for iVent 201 Series
Page 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13...
Page 2: ...VersaMed a General Electric Company doing business as GE Healthcare...
Page 4: ......
Page 14: ......
Page 42: ......
Page 68: ...56 Notes...
Page 70: ......
Page 98: ...86 Notes...
Page 100: ......
Page 140: ...128 Notes...
Page 142: ......
Page 152: ......
Page 154: ......
Page 184: ......
Page 201: ...7 Care Maintenance and Tests 189 Notes...
Page 202: ......
Page 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY...
Page 204: ......
Page 210: ...198 Notes...
Page 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY...
Page 212: ......
Page 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY...
Page 216: ......
Page 219: ...Appendix C Operating Theory 207...
Page 220: ......
Page 221: ...Appendix C Operating Theory 209 Notes...
Page 222: ......
Page 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY...
Page 224: ......
Page 233: ...Appendix D Theory of Breath Delivery 221 Notes...
Page 234: ......
Page 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES...
Page 236: ......
Page 243: ...Appendix E Ventilation Modes 231 Notes...
Page 244: ......
Page 246: ......
Page 252: ...240 Notes...
Page 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES...
Page 254: ......
Page 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR...
Page 278: ......
Page 280: ......
Page 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES...
Page 282: ......
Page 286: ...274...
Page 289: ...Index 277...