166
Warning
Set Count or
Time of
Condition
Mode of Failure Priority Probable Cause Criterion for
Clearing Alarm
Corrective Action
Assess patient
Volume
Limit Reached
4 breaths
delay
Set volume limit
has been
reached in
pressure control
modes
Mid Clinical
situation
Inspiratory
pressure set too
high
Volume limit is
set too low
Limit volume
(V
T
) is not
reached for 2
breaths
Assess patient
Reduce
Inspiratory
pressure
Increase volume
limit
High PEEP
Alarm
2 breath delay
Actual PEEP
value more than
10 cmH
2
O from
set value
High Exhalation
Valve
occlusion
Improper
parameter
setting
PEEP pressure
decrease to set
value + 10 for 1
breath
Check exhalation
valve
Check parameter
setting
Low Minute
Volume
2 breath delay
(20 with
setting
change)
Minute volume
lower than low
alarm setting
High Clinical
Condition
Alarm set too
High
Volume is
higher than the
alarm set for
more than 1
breath
Assess patient
Increase alarm
setting if
appropriate
Decrease trigger
sensitivity
Summary of Contents for iVent 201 1.4
Page 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Page 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Page 4: ......
Page 14: ......
Page 42: ......
Page 68: ...56 Notes ...
Page 70: ......
Page 98: ...86 Notes ...
Page 100: ......
Page 140: ...128 Notes ...
Page 142: ......
Page 152: ......
Page 154: ......
Page 184: ......
Page 201: ...7 Care Maintenance and Tests 189 Notes ...
Page 202: ......
Page 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Page 204: ......
Page 210: ...198 Notes ...
Page 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Page 212: ......
Page 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Page 216: ......
Page 219: ...Appendix C Operating Theory 207 ...
Page 220: ......
Page 221: ...Appendix C Operating Theory 209 Notes ...
Page 222: ......
Page 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Page 224: ......
Page 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Page 234: ......
Page 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Page 236: ......
Page 243: ...Appendix E Ventilation Modes 231 Notes ...
Page 244: ......
Page 246: ......
Page 252: ...240 Notes ...
Page 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Page 254: ......
Page 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Page 278: ......
Page 280: ......
Page 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Page 282: ......
Page 286: ...274 ...
Page 289: ...Index 277 ...