160
Warning
Set Count or
Time of
Condition
Mode of Failure Priority Probable Cause Criterion for
Clearing Alarm
Corrective Action
High O
2
20 breath
delay
Oxygen
concentration
exceeds the high
alarm setting
High
Improper alarm
setting
Change in
patient
breathing
pattern with low
flow adapter
O
2
sensor out of
calibration
FiO
2
is lower
than alarm
setting for 2
breaths
Adjust alarm
setting
Verify O
2
concentration
with external
analyzer
Refer unit to
qualified service
technician for
calibration of O
2
sensor
High Heart
Rate
2 breath delay
The heart rate
exceeds the
upper rate limit
Mid
Clinical Condition Heart rate is
lower than the
alarm settings
for 1 breath
Check the
patient’s
condition.
High Minute
Volume
4 breath delay
(20
with setting
change)
Minute volume
exceeds high
alarm setting
High Clinical
condition
Alarm set too low
Breath rate set
too high
Volume is lower
than alarm
setting, for
more than 2
breaths
Assess patient
Increase alarm
setting if
appropriate
Decrease trigger
sensitivity
Summary of Contents for iVent 201 1.4
Page 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Page 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Page 4: ......
Page 14: ......
Page 42: ......
Page 68: ...56 Notes ...
Page 70: ......
Page 98: ...86 Notes ...
Page 100: ......
Page 140: ...128 Notes ...
Page 142: ......
Page 152: ......
Page 154: ......
Page 184: ......
Page 201: ...7 Care Maintenance and Tests 189 Notes ...
Page 202: ......
Page 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Page 204: ......
Page 210: ...198 Notes ...
Page 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Page 212: ......
Page 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Page 216: ......
Page 219: ...Appendix C Operating Theory 207 ...
Page 220: ......
Page 221: ...Appendix C Operating Theory 209 Notes ...
Page 222: ......
Page 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Page 224: ......
Page 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Page 234: ......
Page 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Page 236: ......
Page 243: ...Appendix E Ventilation Modes 231 Notes ...
Page 244: ......
Page 246: ......
Page 252: ...240 Notes ...
Page 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Page 254: ......
Page 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Page 278: ......
Page 280: ......
Page 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Page 282: ......
Page 286: ...274 ...
Page 289: ...Index 277 ...