1 Introduction
17
O2 SUPPLY SPECIFICATIONS
Parameter
High Pressure
Pressure Range 40 –75psi (2.8-5.1 bar).
Warning: The exact oxygen pressure which the
iVent
TM
201
uses depends on the version you
have. Check the back label (shown in Figure 11)
for the pressure levels of your unit.
Low Pressure
Maximum flow 15L/min or 0.5 psi
VENTILATION PERFORMANCE AND CONTROLLED
PARAMETERS
Parameter
Respiratory Rate
1 – 12 ±1, 12 – 80 ± 2 bpm
Tidal Volume
50-2000 mL
Accuracy of Tidal Volume
Delivery
±10% or ±10 mL whichever is greater
Accuracy of Exhale Tidal
volume Measurement
±15% above 100 mL from actual reading
or ±10 cc below 100 mL
Inspiratory Pressure Limit 5 to 80 ±5 cmH
2
O
Inspiratory Time
Adaptive Time
TM
or 0.2 to 3 ±10% sec
Peak Flow (PIF)
Adaptive Flow
TM
or to 120 ±10% L/min
Spontaneous to 180 ±10% L/min
Oxygen Mix (FiO
2
)
21% to 100% ±10% of the set value or ±3%
absolute, whichever is greater
PEEP
0 to 40 ±1 cmH
2
O or ±10%, whichever is
greater
Trigger Sensitivity
1 to 20 L/min Flow, Off
- 0.5 to – 20 cmH
2
O Pressure, Off
PSV
0 to 60 ±10% cmH
2
O
Positive Pressure Relief
Valve
80 cmH
2
O
Pneumatic Nebulizer
5 to 240 minutes
Controlled Pressure
5 to 80 cmH
2
O ±5 cmH
2
O
FiO
2
at Power up
21%, 40%, 60%, 100% (configurable)
Configurable Humidifier
Setting (Purging Cycle)
HME (10 minute), Heated (1 minutes), Off
Summary of Contents for iVent 201 1.4
Page 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Page 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Page 4: ......
Page 14: ......
Page 42: ......
Page 68: ...56 Notes ...
Page 70: ......
Page 98: ...86 Notes ...
Page 100: ......
Page 140: ...128 Notes ...
Page 142: ......
Page 152: ......
Page 154: ......
Page 184: ......
Page 201: ...7 Care Maintenance and Tests 189 Notes ...
Page 202: ......
Page 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Page 204: ......
Page 210: ...198 Notes ...
Page 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Page 212: ......
Page 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Page 216: ......
Page 219: ...Appendix C Operating Theory 207 ...
Page 220: ......
Page 221: ...Appendix C Operating Theory 209 Notes ...
Page 222: ......
Page 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Page 224: ......
Page 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Page 234: ......
Page 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Page 236: ......
Page 243: ...Appendix E Ventilation Modes 231 Notes ...
Page 244: ......
Page 246: ......
Page 252: ...240 Notes ...
Page 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Page 254: ......
Page 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Page 278: ......
Page 280: ......
Page 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Page 282: ......
Page 286: ...274 ...
Page 289: ...Index 277 ...