6 Alarms
147
Figure 134: The Respiratory Rate Alarm Slider Selected
4.
Press the Control Knob to change the alarm setting for the
Respiratory Rate
.
5.
The top of the slider is selected in bright blue. Turn the dial to adjust the
alarm rate for the upper limit in
Breaths per Minute
. Note that if you
attempt to adjust the upper limit of the alarm rate below the currently-
selected lower limit, you push the lower limit down.
6.
Press the Control Knob to confirm your selection.
7.
The bottom of the slider is selected in bright blue. Turn the dial to adjust the
alarm rate for the lower limit in
Breaths per Minute
. Note that adjusting the
lower limit higher than the currently-selected higher limit moves the higher
limit up.
8.
Press the Control Knob to confirm the alarm settings.
9.
The
Alarm Setting
screen is re-displayed. You can dial and select an
additional
alarm setting to adjust
or dial and select Accept to
accept all the
alarm settings and return to the
Main
screen.
Setting the Minute Volume Alarm
To adjust the minute volume alarms upper and lower limits:
1.
Select the
Menu
option in the
Main
screen to view the
Main Menu
.
2.
Select and choose the first item on the
Main
Menu,
Alarm Settings
.
3.
In the
Alarm Settings
screen, dial the Control Knob until
Minute Volume
is
selected in relief (Figure 135).
Summary of Contents for iVent 201 1.4
Page 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Page 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Page 4: ......
Page 14: ......
Page 42: ......
Page 68: ...56 Notes ...
Page 70: ......
Page 98: ...86 Notes ...
Page 100: ......
Page 140: ...128 Notes ...
Page 142: ......
Page 152: ......
Page 154: ......
Page 184: ......
Page 201: ...7 Care Maintenance and Tests 189 Notes ...
Page 202: ......
Page 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Page 204: ......
Page 210: ...198 Notes ...
Page 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Page 212: ......
Page 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Page 216: ......
Page 219: ...Appendix C Operating Theory 207 ...
Page 220: ......
Page 221: ...Appendix C Operating Theory 209 Notes ...
Page 222: ......
Page 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Page 224: ......
Page 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Page 234: ......
Page 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Page 236: ......
Page 243: ...Appendix E Ventilation Modes 231 Notes ...
Page 244: ......
Page 246: ......
Page 252: ...240 Notes ...
Page 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Page 254: ......
Page 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Page 278: ......
Page 280: ......
Page 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Page 282: ......
Page 286: ...274 ...
Page 289: ...Index 277 ...