6 Alarms
169
THE SENSOR FAILURE ALARM
When the iVent
TM
201 detects a
Sensor Disconnect
alarm or
Check Sensor
condition it enters
Open Loop
ventilation backup mode. During
Open Loop
,
ventilation is provided based on an average of the previous inhaled tidal volumes
(See Appendix E).
If the
Sensor Disconnect
alarm triggers the following message pop-up window
(Figure 149):
Figure 149: the Sensor Disconnect Pop-Up Window
Press the
Clear
button.
Check the connection of flow transducer tubing to the ventilator.
Then, check for excessive amounts of water or secretions, which can set off
the sensor alarm, in the tubing and flow transducer.
If no water or secretions are visible, wait approximately 10 breaths –
sometimes coughing can cause
Check Sensor
or
Sensor Disconnect
alarms.
If the ventilator does not automatically return to the previous mode and
setup, press the Restore option in the Backup Mode Activated message and
check the setting values: PEEP, Rise-Time (auto).
WARNING
If the Check Sensor alarm persists repeatedly, provide an alternative
source of ventilation. Then replace the patient circuit and run the O.V.T.
test.
If the “Patient Disconnect” alarm is available and switched off, the Sensor
Failure alarm and backup mode is not activated.
NOTE
If the
Sensor Disconnect
alarm is detected three (3) times within a minute,
the ventilator does not restore the previous mode automatically, but remains
in Open Loop instead. Press the
Restore
option in the
Backup Mode
Activated
pop-up window to restore the ventilator to the previous mode.
Summary of Contents for iVent 201 1.4
Page 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Page 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Page 4: ......
Page 14: ......
Page 42: ......
Page 68: ...56 Notes ...
Page 70: ......
Page 98: ...86 Notes ...
Page 100: ......
Page 140: ...128 Notes ...
Page 142: ......
Page 152: ......
Page 154: ......
Page 184: ......
Page 201: ...7 Care Maintenance and Tests 189 Notes ...
Page 202: ......
Page 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Page 204: ......
Page 210: ...198 Notes ...
Page 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Page 212: ......
Page 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Page 216: ......
Page 219: ...Appendix C Operating Theory 207 ...
Page 220: ......
Page 221: ...Appendix C Operating Theory 209 Notes ...
Page 222: ......
Page 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Page 224: ......
Page 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Page 234: ......
Page 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Page 236: ......
Page 243: ...Appendix E Ventilation Modes 231 Notes ...
Page 244: ......
Page 246: ......
Page 252: ...240 Notes ...
Page 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Page 254: ......
Page 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Page 278: ......
Page 280: ......
Page 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Page 282: ......
Page 286: ...274 ...
Page 289: ...Index 277 ...