5 Adaptive Bi-Level
139
Exclusion – Re-evaluate the patient's clinical status for causes of high respiratory
drive to exclude premature I:E cycling and manage appropriately.
Intervention – Lower the percent peak flow termination (Esens) and/or decrease
the rise time. Use the waveform package to visualize waveforms and prolong the
I-time until a visually detectable inspiratory flow termination pattern appears
prior to pressure release.
Table 9: Recommended Masks
Part Number
Description
Nasal Masks :
All Types (chin strap may be required for mouth
breathers
Respironics Image 3 SE disposable version
Resmed Mirage 2- reusable version (no vent holes)
M1162031
Koo Medical Equipment- Bluestar Plus NIV Face
Mask Kit,
Ultra Large (KM-310)
M1162032
Koo Medical Equipment- Bluestar Plus NIV Face
Mask Kit,
Large (KM-311)
M1162033
Koo Medical Equipment- Bluestar Plus NIV Face
Mask Kit,
Medium (KM-312)
M1162034
Koo Medical Equipment- Bluestar Plus NIV Face
Mask Kit,
Small (KM-313)
M1162035
Koo Medical Equipment- Bluestar Plus NIV Face
Mask Kit,
Child (KM-314)
Summary of Contents for iVent 201 1.4
Page 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Page 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Page 4: ......
Page 14: ......
Page 42: ......
Page 68: ...56 Notes ...
Page 70: ......
Page 98: ...86 Notes ...
Page 100: ......
Page 140: ...128 Notes ...
Page 142: ......
Page 152: ......
Page 154: ......
Page 184: ......
Page 201: ...7 Care Maintenance and Tests 189 Notes ...
Page 202: ......
Page 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Page 204: ......
Page 210: ...198 Notes ...
Page 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Page 212: ......
Page 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Page 216: ......
Page 219: ...Appendix C Operating Theory 207 ...
Page 220: ......
Page 221: ...Appendix C Operating Theory 209 Notes ...
Page 222: ......
Page 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Page 224: ......
Page 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Page 234: ......
Page 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Page 236: ......
Page 243: ...Appendix E Ventilation Modes 231 Notes ...
Page 244: ......
Page 246: ......
Page 252: ...240 Notes ...
Page 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Page 254: ......
Page 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Page 278: ......
Page 280: ......
Page 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Page 282: ......
Page 286: ...274 ...
Page 289: ...Index 277 ...