Appendix A: Specifications
SpO2 Specifications
A-6
Service Manual
The Masimo SpO
2
parameter has been validated for low perfusion accuracy in bench top testing against a
Biotek Index 2 simulator and Masimo’s simulator with signal strengths of greater than 0.02% and a %
transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or
minus, one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
Note:
Because pulse oximeter equipment measurements are statistically distributed, only about two-
thirds of pulse oximeter equipment measurements can be expected to fall within ± Arms of the
value measured by a COOximeter.
Pulse Rate accuracy has been validated for the range of 25-240 bpm in bench top testing against a Biotek
Index 2 simulator. This variation equals plus or minus one standard deviation which encompasses 68% of
the population.
A.6.4.2 Specifications
Warmer Expected Service Life
The product is designed to meet a life span of 8 years. However,
with proper maintenance and repairs, the service life can be
extended as long as service parts are available.
Measurement Range
•
SpO
2
: 1-100% SpO
2
•
Pulse Rate: 25-240 bpm
Wavelengths
Red
Infrared
660 nm, nominal
905 nm, nominal
Maximum Optical Output Power
Radiant Power at 50 mA pulsed
0.13 mW, minimum
0.79 mW, maximum
Accuracy
•
SpO
2
: from 70-100% SpO
2
, +/- 3% SpO
2
for neonates
•
Below 69%, unspecified
•
Pulse rate, no motion: +/- 3 bpm
•
Pulse rate, motion: +/- 5 bpm
Data Update Period
At maximum averaging time, the data is updated in less than
thirty seconds; shorter averaging times reduce this update
period.
Alarms
User selectable upper and lower limits for SpO
2
and pulse rates
Standards
Designed to ISO 80601-2-61 with the exclusion to clause
201.15.3.5.101.1 with regard to shock and vibration.
Summary of Contents for Giraffe
Page 4: ...RH 2 Service Manual ...
Page 12: ...ii Service Manual About this Manual User Responsibility ...
Page 60: ...Chapter 2 Installation Maintenance and Checkout Scale Checkout Procedures 2 8 Service Manual ...
Page 62: ...Chapter 3 Calibration Scale Calibration Bedded Models Only 3 2 Service Manual ...
Page 172: ...Chapter 6 Illustrated Parts Wiring Diagrams 6 40 Service Manual ...