Appendix B: Electromagnetic Compatibility
Manufacturer’s Guidance and Declaration Regarding Electromagnetic Immunity
B-2
Service Manual
B.3 Manufacturer’s Guidance and Declaration Regarding Electromagnetic
Immunity
The Giraffe/Panda Warmer is intended for use in the electronic environment specified below. The user of the
Giraffe/Panda Warmer should ensure that it is used in such an environment.
Note:
Ut is the main voltage before application of the test level.
Electromagnetic Immunity
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic Environment
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material the relative humidity
should be at least 30%.
Connectors marked with ESD symbol
excluded.
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power
supply line.
± 1kV for input/
output line.
± 2kV for power supply
line.
± 1kV for input/output
line.
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1kV differential
Mode.
± 2kV common
mode line.
± 1kV differential Mode.
± 2kV common mode
line.
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5 % Ut (> 95 %
dip in Ut) for
0.5 cycle
40 % Ut (60 %
dip in Ut) for
5 cycles
70% Ut (30 % dip
in Ut) for 25
cycles
< 5% Ut (> 95 %
dip in Ut) for
5 sec.
< 5 % Ut (> 95 % dip in
Ut) for 0.5 cycle
40 % Ut (60 % dip in Ut)
for 5 cycles
70 % Ut (30 % dip in Ut)
for 25 cycles
< 5 % Ut (> 95 % dip in
Ut) for 5 sec.
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Giraffe/
Panda Warmer requ continued
operation during power mains
interruptions, it is recommended that the
Giraffe/Panda Warmer be powered from
an uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
environment
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital.
Summary of Contents for Giraffe
Page 4: ...RH 2 Service Manual ...
Page 12: ...ii Service Manual About this Manual User Responsibility ...
Page 60: ...Chapter 2 Installation Maintenance and Checkout Scale Checkout Procedures 2 8 Service Manual ...
Page 62: ...Chapter 3 Calibration Scale Calibration Bedded Models Only 3 2 Service Manual ...
Page 172: ...Chapter 6 Illustrated Parts Wiring Diagrams 6 40 Service Manual ...
Page 180: ...Appendix A Specifications SpO2 Specifications A 8 Service Manual ...