Chapter 1: Functional Description
RS-232 Serial Data
Service Manual
1-27
1.4 RS-232 Serial Data
Note:
In the event of a power failure, all serial communication will cease until power is restored.
1.4.1 RS-232 Connector
The Nurse Call and the serial data output share the same female, nine pin, d-type connector.
•
Pin 2: Receive Data (incubator input)
•
Pin 3: Transmit Data (incubator output)
•
Pin 5: Gnd (Signal Ground)
1.4.1.1 Cable Requirements
The user interface cable must have capacitance less than 1500 pF. It should be a shielded cable such as
Belden 9611 with AMP shielding kit 748046-1 and ferrule 747579-8.
1.4.2 Data Stream
The data stream from the Panda and Giraffe products is repeated every two seconds. The RS-232
parameters are 116000 bps baud rate, no parity, 8 data bits, and one stop bit. The data is in ASCII format;
the string is described below:
PWRM_2.0.0,36.8,36.4,0,P,0,0,0,00,29
Warning:
The computer or RS-232 monitor’s user program must continuously check the data
link. The program should constantly verify connection to the control panel and check for
updated data.
HYB
This is the product code.
•
GWRM
= Giraffe Warmer
•
PWRM
= Panda Warmer
•
PNBW
= Panda Non-Bedded Warmer
2.0.0
This is the control board software revision (thermoregulation in the startup screen).
36.8
Temperature from Patient Probe #1.
OPEN
Temperature from Patient Probe #2.
OPEN
represents no probe connected.
36.4
Baby (Servo) mode desired temperature setting.
0
Heater power.
P
•
Baby (Servo) Mode = P
•
Manual Mode = M
Summary of Contents for Giraffe
Page 4: ...RH 2 Service Manual ...
Page 12: ...ii Service Manual About this Manual User Responsibility ...
Page 60: ...Chapter 2 Installation Maintenance and Checkout Scale Checkout Procedures 2 8 Service Manual ...
Page 62: ...Chapter 3 Calibration Scale Calibration Bedded Models Only 3 2 Service Manual ...
Page 172: ...Chapter 6 Illustrated Parts Wiring Diagrams 6 40 Service Manual ...
Page 180: ...Appendix A Specifications SpO2 Specifications A 8 Service Manual ...