Chapter 4: Troubleshooting
PC Command Line Interface
4-16
Service Manual
4.2.3.4 logs
.
4.2.3.5 pulsetone
command: pulsetone
arguments:
‘vol’ = desired speaker volume (optional)
1(min) - 4(max)
function:
Sets the volume to the specified level and generates a pulsetone.
If the argument ‘vol’ is not specified, a pulsetone will be generated at the current volume setting.
usage: pulsetone [vol]
4.2.3.6 ncas
Note:
NCAS is an early acronym for Hands Free Alarm Silence.
command: ncas
arguments: none
function:
Retrieves and displays the instantaneous state of the Hands Free Alarm Silence input to the main
processor. The state is reported as ‘active’ or ‘inactive’.
usage: ncas
4.2.3.7 dips
command: dips
arguments: none
function: Reads and prints out the status of the control board DIP switches.
usage: dips
4.2.3.8 heateng
command: heateng
arguments:
‘info’ = report sw version and heater/board eeprom id’s
‘stat’ = report operational state, safety relay and tstat states and heater output power