3
Operating instructions
46
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
1
use the
right
and
left
arrow keys to position the cursor in the
temperature
field, and
2
use the
up
and
down
arrow keys to adjust the value.
It may take up to 30 minutes to reach the set temperature, depending on
ambient conditions.
Isoelectric focusing will begin after the rehydration step is complete. The HV ON
lamp will light and the
status
screen showing the step currently in progress as
well as all parameters will appear:
Note:
The indicated wattage is the total power delivered to all strips, and will
often read 0 W because any value less than 0.5 is rounded down to 0. If the
delivered power exceeds 0.4 W per strip, stop the run and reduce the V or
µA settings.
The current platform temperature is indicated.
The average current per IPG strip (µA), the total instantaneous power delivered
(W), and actual voltage (V) are indicated. If the current is the limiting parameter,
the “A” in “µA” will flash.
The Ettan IPGphor 3 will display the volt-hours during the run as:
•
VhS, which indicates the total volt hours for that step, or
•
VhT, the total volt-hours accumulated during the entire run at the current
time.
Hours elapsed for the current step will be indicated by Hrs.
R u n n i n g P r o t # 1 @ 2 0
˚
C
S t e p 1 o f 3 V h r s
S t e p V h r s E l a p s e d
5 0 0 V 4 0
µ
A / S t r i p 0 . 0 W
Summary of Contents for Ettan IPGphor 3
Page 1: ...GE Healthcare Ettan IPGphor 3 Instrument User Manual Ettan...
Page 2: ......
Page 4: ......
Page 54: ...4 Care and maintenance 54 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Page 62: ...6 Customer service information 62 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Page 63: ......