Operating instructions
40
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
The original factory programmed IEF temperature is 20 °C. To adjust the
temperature
:
1
use the
left
and
right
arrow keys to position the cursor in the active °C field,
and
2
use the
up
and
down
arrow keys to select the value within the range of 15–
30 °C.
The original factory programmed value is 50 µA per IPG strip.
To adjust the
current limit
:
1
use the
left
and
right
arrow keys to position the cursor under one of the
three active µA digits, and
2
use the
up
and
down
arrow keys to select the value within the range
0–200 µA per strip.
Note:
The total deliverable amperage is 1500 µA.
3
Use the arrow keys to adjust the numbers of strips to run.
4
To advance to the next edit screen, press the
EDIT
key.
IEF programming
The second screen in edit mode is the first in a series of nine programmable IEF
steps in which the voltage change pattern, voltage, and step duration are set. The
steps are displayed in groups of 3 steps per screen. The number of steps which
have been programmed is displayed next to the protocol number.
Each step is identified by “S” and the step number and can be accessed by
placing the cursor under the step number using the
right
and
left
arrow keys. To
advance to the next step group press the
down
arrow key and to return to a
previous step group press the
up
arrow key.
P r o t # 1 F i l e 1
R e h y d r a t e 1 0 : 0 0 H r @ 2 0
˚
C
I E F @ 2 0
˚
C 5 0
µ
A / S t r i p
1 S t r i p s 0 S t e p s
P r o t # 1 F i l e 1
R e h y d r a t e 1 0 : 0 0 H r @ 2 0
˚
C
I E F @ 2 0
˚
C 5 0
µ
A / S t r i p
1 S t r i p s 0 S t e p s
Summary of Contents for Ettan IPGphor 3
Page 1: ...GE Healthcare Ettan IPGphor 3 Instrument User Manual Ettan...
Page 2: ......
Page 4: ......
Page 54: ...4 Care and maintenance 54 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Page 62: ...6 Customer service information 62 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Page 63: ......